Fujikado Takashi, Kuroda Teruhito, Maeda Naoyuki, Ninomiya Sayuri, Goto Hiroya, Tano Yasuo, Oshika Tetsuro, Hirohara Yoko, Mihashi Toshifumi
Ophthalmology, Osaka University Graduate School of Medicine, Osaka, Japan.
J Cataract Refract Surg. 2004 Jun;30(6):1198-208. doi: 10.1016/j.jcrs.2003.12.023.
To predict the visual deterioration of eyes with cortical (CC) or nuclear (NC) cataract from objective data on ocular higher-order aberration (HOA) and forward (FLS) and backward light scattering (BLS).
Osaka University Medical School, Osaka, Japan.
Twenty-two eyes with mild NC, 41 eyes with mild CC, and 11 normal eyes were examined. Higher-order aberrations were calculated with the Zernike polynomials up to the fourth order from the values obtained by wavefront analysis using the Hartmann-Shack aberrometer. Forward light scattering was calculated from the size of the aberrometer spot images for the central 4 mm, and backward light scattering (BLS) was calculated from the optical density of the Scheimpflug images. The relationship between the area under the log contrast sensitivity function (AULCSF) curve and HOAs, FLS, and BLS was examined.
Area under the log contrast sensitivity function was moderately correlated with the HOAs, FLS, and BLS. Multiple linear regression analysis revealed that the AULCSF was predicted by the linear combination of these variables (R(2)=.484, P<.001). Area under the log contrast sensitivity was predicted by BLS and HOA (R(2)=.555) in the NC group and by FLS and HOAs (R(2)=.540) in the CC group.
Loss of contrast sensitivity was predominantly due to BLS and HOA in eyes with NC and FLS and HOA in eyes with CC. Higher-order aberrations, FLS, and BLS, variables that are obtained objectively, can be used to predict quantitatively the visual deterioration in cataractous eyes.
根据眼部高阶像差(HOA)、前向(FLS)和后向光散射(BLS)的客观数据,预测皮质性(CC)或核性(NC)白内障患者眼睛的视力恶化情况。
日本大阪大学医学院。
对22只轻度核性白内障眼、41只轻度皮质性白内障眼和11只正常眼进行检查。使用Hartmann-Shack像差仪通过波前分析获得的值,用Zernike多项式计算直至四阶的高阶像差。根据中央4mm像差仪光斑图像的大小计算前向光散射,并根据Scheimpflug图像的光密度计算后向光散射(BLS)。研究对数对比敏感度函数(AULCSF)曲线下面积与HOA、FLS和BLS之间的关系。
对数对比敏感度函数曲线下面积与HOA、FLS和BLS呈中度相关。多元线性回归分析显示,这些变量的线性组合可预测AULCSF(R² = 0.484,P < 0.001)。核性白内障组中,对数对比敏感度曲线下面积由BLS和HOA预测(R² = 0.555);皮质性白内障组中,由FLS和HOA预测(R² = 0.540)。
核性白内障眼的对比敏感度下降主要归因于BLS和HOA,皮质性白内障眼则归因于FLS和HOA。高阶像差、FLS和BLS这些客观获得的变量可用于定量预测白内障眼的视力恶化情况。