人体姿势改变和运动过程中心输出量的无创评估:Modelflow与脉搏染料密度测定法的比较

Noninvasive evaluation of cardiac output during postural change and exercise in humans: comparison between the modelflow and pulse dye-densitometry.

作者信息

Matsukawa K, Kobayashi T, Nakamoto T, Murata J, Komine H, Noso M

机构信息

Department of Physiology, Graduate School of Health Sciences, Hiroshima University, Hiroshima, 734-8551 Japan.

出版信息

Jpn J Physiol. 2004 Apr;54(2):153-60. doi: 10.2170/jjphysiol.54.153.

Abstract

To investigate whether the Model-flow method, by simulating the aortic input impedance model from a noninvasive monitoring of arterial blood pressure, reflected a reliable measure of cardiac output (CO) during postural change and whole-body exercise occurring in daily life, we compared the Modelflow-estimated CO with a simultaneous reference determined by the pulse dye-densitometry. Nine healthy volunteers performed postural change from supine to upright and dynamic stepping exercise. The Modelflow-estimated CO decreased to 4.8 +/- 0.5 l/min, from 5.8 +/- 0.6 l/min, during the postural change and increased to 12.8 +/- 1.3 l/min during a stepping exercise, returning to 5.1 +/- 0.4 l/min at 5 min after exercise. When comparing the pooled data of CO during resting and following exercise between the Modelflow and pulse dye-densitometry, we found that the average CO did not differ between the two estimates and that there was a significant correlation between them; the slope of the linear regression line corresponded to approximately 1.0. Although such linear relationship was also observed in an individual subject, the slope of the regression line varied from 0.737 to 1.588 among the subjects. The calibration of the Modelflow-estimated CO with the dye-densitometry value at supine or upright improved a correlation between the two estimates. Thus it is likely that the noninvasive Modelflow simulation from arterial blood pressure can provide a reliable estimation of group-average cardiac output during postural change and stepping exercise occurring in daily life. It will be recommended for a more accurate estimation of cardiac output in a given subject to calibrate the Modelflow data with an independent measure.

摘要

为了研究通过对动脉血压进行无创监测来模拟主动脉输入阻抗模型的模型流方法,是否能反映日常生活中姿势变化和全身运动期间的心输出量(CO)的可靠测量值,我们将模型流估计的CO与通过脉搏染料密度测定法同时确定的参考值进行了比较。九名健康志愿者进行了从仰卧到直立的姿势变化和动态踏步运动。在姿势变化期间,模型流估计的CO从5.8±0.6升/分钟降至4.8±0.5升/分钟,在踏步运动期间增加至12.8±1.3升/分钟,运动后5分钟恢复至5.1±0.4升/分钟。当比较模型流和脉搏染料密度测定法在静息和运动后CO的汇总数据时,我们发现两种估计值之间的平均CO没有差异,并且它们之间存在显著相关性;线性回归线的斜率约为1.0。尽管在个体受试者中也观察到了这种线性关系,但受试者之间回归线的斜率在0.737至1.588之间变化。用仰卧或直立时的染料密度测定值对模型流估计的CO进行校准,改善了两种估计值之间的相关性。因此,通过动脉血压进行的无创模型流模拟可能能够可靠地估计日常生活中姿势变化和踏步运动期间的群体平均心输出量。建议使用独立测量方法对模型流数据进行校准,以便更准确地估计特定受试者的心输出量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验