Astier Y, Bartlett P N
School of Chemistry, University of Southampton, Southampton SO17 1BJ, UK.
Bioelectrochemistry. 2004 Aug;64(1):53-9. doi: 10.1016/j.bioelechem.2004.01.004.
We report a new approach to the measurement of alkaline phosphatase concentration based on the use of a disposable poly(aniline) microelectrochemical transistor. The measurement is carried out in a two cell configuration in which the poly(aniline) microelectrochemical transistor operates in acid solution and is connected to the alkaline buffer solution containing the analyte by a salt bridge. Disposable microelectrochemical transistors were reproducibly fabricated by electrochemical deposition of poly(aniline) onto photolithographically fabricated gold microband arrays. Using these devices alkaline phosphatase was detected by employing p-aminophenyl phosphate as the substrate for the enzyme and using glucose and glucose oxidase to recycle the p-aminophenol generated upon enzyme catalysed hydrolysis of the phosphate. Recycling the p-aminophenol with glucose and glucose oxidase amplified the detection of alkaline phosphatase approximately tenfold. Using this approach we obtain linear calibration curves for alkaline phosphatase up to 5 nM within 70 s on single use devices.