Biomechanical study of lumbar pedicle screws in a corpectomy model assessing significance of screw height.

作者信息

Mikles Mark R, Asghar Ferhan A, Frankenburg Elizabeth P, Scott Derek S, Graziano Greg P

机构信息

Department of Orthopaedics, University of Michigan, Ann Arbor, MI 48109-0328, USA.

出版信息

J Spinal Disord Tech. 2004 Aug;17(4):272-6. doi: 10.1097/01.bsd.0000095400.27687.89.

Abstract

BACKGROUND

We tested the hypothesis that a pedicle screw construct's height is an important factor in strengthening a screw-rod system.

METHODS

Six corpectomy constructs were made, each using two ultra-high-molecular-weight polyethylene blocks, 6.5-mm pedicle screws, and two 6.35-mm rods. Pedicle screws were placed at +10-, +5-, 0-, and -5-mm depths in relation to the dorsal surface of the corpectomy model. Nondestructive testing was performed in flexion/extension and in torsion.

RESULTS

For all modes tested, the screw-rod constructs continued to increase in stiffness as the height of the construct was lowered, and this was statistically significant at all heights tested (P < 0.001). The stiffness increased 232% when comparing flexion at +10 and -5 mm and increased 231% in extension from +10 to -5 mm. The torsional stiffness increased 171% when comparing +10 and -5 mm.

CONCLUSIONS

Thus, lower-profile instrumentation systems should be used to take advantage of this by decreasing the size and bulkiness of the implants while increasing the strength of the construct.