高压离子阱质谱分析法

High-pressure ion trap mass spectrometry.

作者信息

Whitten William B, Reilly Peter T A, Ramsey J Michael

机构信息

Oak Ridge National Laboratory, PO Box 2008 MS 6142, Oak Ridge, TN 37831-6142, USA.

出版信息

Rapid Commun Mass Spectrom. 2004;18(15):1749-52. doi: 10.1002/rcm.1549.

Abstract

The effects of buffer gas pressure on ion trap stability, mass resolution/calibration, and choice of mass scanning are described. Pressure effects were treated phenomenologically by adding a drag term to the ion equations of motion. The resulting collisional damping enlarges the mass-dependent stability region but reduces the region in which mass-selective resonance ejection can be performed. The pressure effects can be reduced by increasing the frequency of the alternating quadrupole field.

摘要

描述了缓冲气体压力对离子阱稳定性、质量分辨率/校准以及质量扫描选择的影响。通过在离子运动方程中添加一个拖曳项,从现象学角度处理了压力效应。由此产生的碰撞阻尼扩大了质量相关的稳定区域,但减小了可以进行质量选择性共振喷射的区域。可以通过提高交变四极场的频率来降低压力效应。