Lam Andrew K C, Lam Chin-hang
Department of Optometry & Radiography, The Hong Kong Polytechnic University, Kowloon, Hong Kong SAR, China.
Optom Vis Sci. 2004 Aug;81(8):597-600. doi: 10.1097/01.opx.0000141795.95597.98.
The Valsalva maneuver is known to affect intraocular pressure (IOP). Simple breath-holding may cause IOP elevation. A recent study demonstrated a decrease in pulsatile ocular blood flow (POBF) during forcible exhalation. This study investigated whether breath-holding could affect POBF measurement.
Thirty-eight healthy young adults were recruited. Their POBF's were measured with an ocular blood flow pneumatonometer. The first set of measurements was made using normal measurement protocol. Three consecutive readings were obtained, and the mean was used for analysis. The second set of measurements was taken after 5-min rest, and the subjects were required to hold their breath during the acquisition period.
As previously reported, POBF reduced with increasing myopia. There was no significant change in IOP, pulse rate, POBF, and pulse amplitude between normal protocol and breath-holding condition. POBF and pulse amplitude demonstrated a greater variation, shown by coefficient of variation, when subjects held their breath.
This study found a greater variation in consecutive POBF measurements during breath-holding condition. There was no significant difference in either POBF or pulse amplitude during breath-holding session, probably because of the use of three consecutive measurements, and averaged results were generated from them. It is advised to measure the POBF by taking consecutive readings, and subjects should not hold their breath.
已知瓦尔萨尔瓦动作会影响眼压(IOP)。简单的屏气可能会导致眼压升高。最近的一项研究表明,用力呼气时搏动性眼血流量(POBF)会减少。本研究调查了屏气是否会影响POBF测量。
招募了38名健康的年轻成年人。使用眼血流气压计测量他们的POBF。第一组测量采用正常测量方案。获得连续三次读数,并将平均值用于分析。第二组测量在休息5分钟后进行,受试者在采集期间需要屏气。
如先前报道,POBF随着近视程度的增加而降低。正常方案和屏气状态之间的眼压、脉搏率、POBF和脉搏幅度均无显著变化。当受试者屏气时,POBF和脉搏幅度的变异系数显示出更大的变异性。
本研究发现屏气状态下连续POBF测量的变异性更大。屏气期间POBF或脉搏幅度均无显著差异,这可能是因为使用了连续三次测量,并从中得出了平均结果。建议通过连续读数来测量POBF,并且受试者不应屏气。