Ernest Paul H
lcon Research, Ltd., Fort Worth, TX, USA.
J Cataract Refract Surg. 2004 Aug;30(8):1755-8. doi: 10.1016/j.jcrs.2003.12.054.
To compare the light transmittance of 4 currently marketed intraocular lenses (IOLs) with that of the new AcrySof Natural IOL (Alcon Laboratories, Inc.), which is designed to mimic the light-attenuating characteristics of the human crystalline lens.
Research laboratory, Alcon Research, Ltd., Fort Worth, Texas, USA.
Light-transmission spectra of 4 commonly implanted IOLs were compared with the spectrum of the AcrySof Natural IOL (model SN60AT).
While the 4 other foldable IOLs transmitted nearly 100% (near 0% absorption) of the light in the 400 to 500 nm (blue-light) region, the AcrySof Natural IOL absorbed a significant portion of blue light. Comparison of the spectrum of the AcrySof Natural IOL with the spectrum of human crystalline lenses of various ages showed that the AcrySof Natural lens closely mimics the light-attenuating characteristics of the human crystalline lens.
The AcrySof Natural IOL provides a transmission spectrum more like that of a human crystalline lens than do other commonly implanted foldable IOLs. The Natural IOL reduces ultraviolet and blue-light exposure to the retina, mimicking the filtering normally provided by the human crystalline lens.
比较4种目前市场上销售的人工晶状体(IOL)与新型AcrySof Natural人工晶状体(爱尔康实验室公司)的透光率,后者旨在模拟人晶状体的光衰减特性。
美国得克萨斯州沃思堡爱尔康研究有限公司研究实验室。
将4种常用植入式人工晶状体的透光光谱与AcrySof Natural人工晶状体(型号SN60AT)的光谱进行比较。
其他4种可折叠人工晶状体在400至500纳米(蓝光)区域传输近100%(吸收近0%)的光,而AcrySof Natural人工晶状体吸收了相当一部分蓝光。将AcrySof Natural人工晶状体的光谱与不同年龄的人晶状体光谱进行比较表明,AcrySof Natural晶状体紧密模拟了人晶状体的光衰减特性。
与其他常用的可折叠人工晶状体相比,AcrySof Natural人工晶状体提供的透射光谱更类似于人晶状体。Natural人工晶状体减少了视网膜受到的紫外线和蓝光照射,模拟了人晶状体通常提供的过滤作用。