采用流动注射分析-荧光检测法测定呼出气冷凝物中的过氧化氢。

Determination of hydrogen peroxide in exhaled breath condensate by flow injection analysis with fluorescence detection.

作者信息

Svensson Sophie, Olin Anna-Carin, Lärstad Mona, Ljungkvist Göran, Torén Kjell

机构信息

Occupational and Environmental Medicine, Sahlgrenska University Hospital and Academy at Göteborg University, Sankt Sigfridsgatan 85, SE-41266 Göteborg, Sweden.

出版信息

J Chromatogr B Analyt Technol Biomed Life Sci. 2004 Oct 5;809(2):199-203. doi: 10.1016/j.jchromb.2004.06.027.

Abstract

A method for the determination of hydrogen peroxide in exhaled breath condensate (EBC) by automated flow injection analysis (FIA) with fluorescence detection was developed and validated. In the enzymatic assay a fluorescent dimer of para-hydroxyphenyl acetic acid (HPAA) was formed by the redox coupling reaction between hydrogen peroxide and horseradish peroxidase (HRP). The calibration curve of hydrogen peroxide was linear over a range of 40-5000 nM. The coefficient of variation (CV) for within-day precision was 1-3%; for between-day precision, it was 2-5% over the validated range. The assay requires a small sample aliquot (150 microl) and no incubation time, and has an analytical runtime of < 2 min. It is therefore suitable for larger studies. The method was used to detect hydrogen peroxide in EBC of asthmatic patients and healthy volunteers. A statistically significant difference was found between patients with asthma (n = 19) and control subjects without asthma (n = 19), 780 nM versus 480 nM (P = 0.03).

摘要

开发并验证了一种通过自动流动注射分析(FIA)结合荧光检测来测定呼出气冷凝物(EBC)中过氧化氢的方法。在酶促测定中,过氧化氢与辣根过氧化物酶(HRP)之间的氧化还原偶联反应形成了对羟基苯乙酸(HPAA)的荧光二聚体。过氧化氢的校准曲线在40 - 5000 nM范围内呈线性。日内精密度的变异系数(CV)为1 - 3%;在验证范围内,日间精密度的变异系数为2 - 5%。该测定法所需样品量小(150微升)且无需孵育时间,分析运行时间< 2分钟。因此适用于更大规模的研究。该方法用于检测哮喘患者和健康志愿者的EBC中的过氧化氢。发现哮喘患者(n = 19)和无哮喘对照受试者(n = 19)之间存在统计学显著差异,分别为780 nM和480 nM(P = 0.03)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验