Suppr超能文献

龈沟液电子体积定量的关键步骤:蒸发、液体潴留、局部条件及重复测量的潜在影响

Critical steps in electronic volume quantification of gingival crevicular fluid: the potential impact of evaporation, fluid retention, local conditions and repeated measurements.

作者信息

Tözüm Tolga F, Hatipoğlu Hasan, Yamalik Nermin, Gürsel Mihtikar, Alptekin Nilgün O, Ataoğlu Tamer, Marakoğlu Ismail, Gürsoy Ulvi K, Eratalay Kenan

机构信息

Department of Periodontology, Faculty of Dentistry, Hacettepe University, Ankara, Turkey.

出版信息

J Periodontal Res. 2004 Oct;39(5):344-57. doi: 10.1111/j.1600-0765.2004.00758.x.

Abstract

BACKGROUND AND OBJECTIVES

Various methodological factors may operate during clinical gingival crevicular fluid (GCF) sampling, volume quantification or subsequent laboratory analysis. For precise volume quantification, specific concern for generation and maintenance of a reliable calibration curve, the potential risk of GCF loss as a result of evaporation or fluid retention on actual volume and the impact of local conditions is needed because each of these factors may act as a source of subsequent volumetric distortions. Thus, the present study aimed to analyse the impact of sample transfer time on the rate of evaporation and the possibility of fluid retention, and the impact of local conditions and number of replicated measurements on the reliability of calibration data.

MATERIALS AND METHODS

To analyse evaporative errors, standardized Periopaper strips provided with known test volumes (0.1 microl, 0.2 microl, 0.5 microl and 0.6 microl) were transferred to Periotron 8000 with different time intervals (immediately, 5 s, 30 s and 60 s). For fluid retention, after quantifying the actual volume of the strips provided with known volumes (0.1 microl and 0.6 microl) of two test fluids, a second set of measurements was performed using dry strips. To determine the impact of local conditions (temperature and humidity) and the validity of 3, 5 and 20 replications (0.0-0.6 microl with 0.1- microl increments) on device calibration for 20 degrees C and 25 degrees C, electronic readings were obtained from three devices at three different locations. Differences in volumetric data in each experimental design were statistically analysed.

RESULTS

No significant fluid loss was observed within 5 s, but evaporation clearly led to volumetric distortions with extending transfer times (30 s or 60 s) (p < 0.05). Measurable amounts of fluid retention were found for both volumes and both test fluids, but not with identical patterns. Local conditions resulted in unique calibration data for each test volume and for each device. Although a 5 degrees C increase generally provided higher readings, this was not observed for all devices at all volumes. Additional replicates (n = 5 or n = 20) did not seem to add any further reliability to the triplicate scores for the given test volumes.

CONCLUSION

The findings of the present study confirm the reliability of triplicate readings, and uniqueness of each device and electronic data and the distinct impact of local environmental conditions on the generation/maintenance of calibration scores for each particular device. Furthermore, they underline time-dependent evaporation and fluid retention as additional technical concerns and once again highlight the importance of methodological standardization of the electronic volume quantification process.

摘要

背景与目的

在临床龈沟液(GCF)采样、体积定量或后续实验室分析过程中,可能存在多种方法学因素的影响。为实现精确的体积定量,需要特别关注可靠校准曲线的生成与维持、由于蒸发或液体残留对实际体积造成的GCF损失风险以及局部条件的影响,因为这些因素中的每一个都可能成为后续体积失真的来源。因此,本研究旨在分析样本转移时间对蒸发速率和液体残留可能性的影响,以及局部条件和重复测量次数对校准数据可靠性的影响。

材料与方法

为分析蒸发误差,将配备已知测试体积(0.1微升、0.2微升、0.5微升和0.6微升)的标准化Periopaper试纸条在不同时间间隔(立即、5秒、30秒和60秒)转移至Periotron 8000。为研究液体残留情况,在对配备已知体积(0.1微升和0.6微升)两种测试液的试纸条实际体积进行定量后,使用干燥试纸条进行第二轮测量。为确定局部条件(温度和湿度)以及3次、5次和20次重复测量(0.0 - 0.6微升,以0.1微升递增)对20摄氏度和25摄氏度下设备校准的有效性,在三个不同位置从三台设备获取电子读数。对每个实验设计中的体积数据差异进行统计学分析。

结果

在5秒内未观察到明显的液体损失,但随着转移时间延长(30秒或60秒),蒸发明显导致体积失真(p < 0.05)。对于两种体积的试纸条和两种测试液均发现有可测量的液体残留量,但模式并不相同。局部条件导致每个测试体积和每个设备都有独特的校准数据。尽管温度升高5摄氏度通常会使读数更高,但并非所有设备在所有体积下都是如此。对于给定的测试体积,额外的重复测量(n = 5或n = 20)似乎并未为三次重复测量得分增加更多可靠性。

结论

本研究结果证实了三次重复读数的可靠性、每个设备和电子数据的独特性以及局部环境条件对每个特定设备校准得分生成/维持的显著影响。此外,研究强调了时间依赖性蒸发和液体残留是额外的技术问题,并再次突出了电子体积定量过程方法标准化的重要性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验