Suppr超能文献

三种液体用于校准新型 Periotron® 8010 的比较。

Comparison of three fluids for calibration of the new Periotron® 8010.

机构信息

Department of Stomatology University of Valencia Gascó i Oliag 1, 46010 Valencia, Spain

出版信息

Med Oral Patol Oral Cir Bucal. 2023 Nov 1;28(6):e519-e524. doi: 10.4317/medoral.25917.

Abstract

BACKGROUND

The aim of the present study was to calibrate the Periotron® model 8010 with volumes of three different fluids (distilled water, serum, and saliva) and to identify which of the three is the most reliable, feasible, and reproducible for routine calibration.

MATERIAL AND METHODS

A total of 450 samples of Periopaper® were divided into three groups (150 each per group): distilled water, serum matrix and saliva. A calibration curve was run with 0.25, 0.50, 0.75, 1.00 and 1.25 µl of each of the fluids, and the results were determined in Periotron units (PU). Statistical analysis was performed with one-way ANOVA followed by Bonferroni's post hoc test and a linear equation.

RESULTS

Distilled water presented the lowest levels of PU at all volumes, while serum showed the highest levels at high volumes. Linear regression equations rendered similar slopes for saliva and distilled water, while serum was statistically different. Saliva presented a reproduction percentage of 99.7%, which indicated better accuracy and precision than serum and distilled water.

CONCLUSIONS

Saliva is more reliable and accurate than water or serum for the purpose of calibration of the Periotron® model 8010, though it shares drawbacks with serum. Distilled water is more easily available and does not require any additional procedure, in addition to producing a similar slope to saliva and a smaller deviation from the media than serum.

摘要

背景

本研究旨在使用三种不同液体(蒸馏水、血清和唾液)对 Periotron® 8010 模型进行校准,并确定哪一种液体最可靠、最可行且最适合常规校准。

材料和方法

共收集了 450 份 Periopaper®样本,分为三组(每组 150 份):蒸馏水组、血清基质组和唾液组。对每种液体的 0.25、0.50、0.75、1.00 和 1.25 µl 进行校准曲线运行,并以 Periotron 单位(PU)确定结果。采用单因素方差分析(ANOVA),然后进行 Bonferroni 事后检验和线性方程分析。

结果

蒸馏水在所有体积下的 PU 值最低,而血清在高体积下的 PU 值最高。唾液和蒸馏水的线性回归方程具有相似的斜率,而血清则具有统计学差异。唾液的重现百分比为 99.7%,表明其准确性和精密度优于血清和蒸馏水。

结论

与水或血清相比,唾液更适合用于校准 Periotron® 8010 模型,因为它更可靠且准确,尽管它与血清一样存在一些缺点。蒸馏水更容易获得,且不需要任何额外的处理程序,并且与唾液相比,其斜率相似,与血清相比,偏差更小。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0b51/10635629/576742657ca4/medoral-28-e519-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验