Intraobserver variability in optical coherence tomography.

作者信息

Browning David J, Fraser Christina M

机构信息

Charlotte Eye, Ear, Nose, and Throat Associates, Charlotte, NC 28210, USA.

出版信息

Am J Ophthalmol. 2004 Sep;138(3):477-9. doi: 10.1016/j.ajo.2004.03.014.

Abstract

PURPOSE

To assess the intraobserver variability of optical coherence tomography (OCT) in patients with clinically stable maculas.

DESIGN

Retrospective, observational case series using a diagnostic instrument.

METHODS

Retrospective study. setting. Private retina practice. patient population. Twenty-two eyes of 19 patients, each studied twice within 4 months. observation procedure. Optical coherence tomograph measurements of the macula obtained by two photographers. main outcome measures. Ordinary least products (OLP) and Bland-Altman analysis of OCT data.

RESULTS

No fixed or proportional bias was detected in foveal zone thickness and total macular volume between measurements of either of the OCT operators. The coefficient of repeatability for foveal zone thickness was 37.0 microm for photographer 1 and 34.8 microm for photographer 2. The coefficient of repeatability for total macular volume was 0.29 mm(3) for photographer 1 and 0.10 mm(3) for photographer 2.

CONCLUSION

Foveal zone thickness and total macular volume measurements show low intraobserver variability when analyzed by OLP and Bland-Altman techniques. Bounds are given for foveal zone thickness and total macular volume for which longitudinal OCT measurements by the same operator can be considered to reflect natural history or intervention effects rather than intraobserver variability.