使用 7 种光学相干断层扫描仪测量健康眼的黄斑厚度在操作者间和操作者内的可重复性。

Macular thickness interoperator and intraoperator reproducibility in healthy eyes using 7 optical coherence tomography instruments.

机构信息

Department of Ophthalmology, University Vita-Salute, Scientific Institute San Raffaele, Via Olgettina 60, Milan, Italy.

出版信息

Am J Ophthalmol. 2010 Aug;150(2):199-204.e1. doi: 10.1016/j.ajo.2010.03.015. Epub 2010 Jun 8.

Abstract

PURPOSE

To evaluate macular thickness measurement reproducibility using 6 new spectral-domain (SD) optical coherence tomography (OCT) devices and 1 time-domain OCT device

DESIGN

Prospective, observational study.

METHODS

setting: Clinical practice. study population: Macular thickness was assessed in 18 randomly chosen consecutive eyes of 18 healthy volunteers by 2 masked operators using 6 SD OCT devices and 1 time-domain OCT device. main outcome measures: Intraoperator and interoperator reproducibility of OCT-measured central macular thickness.

RESULTS

Mean macular thickness ranged from 172.77 to 272.87 microm. Intraobserver Intraclass coefficient correlation ranged from 0.75 to 0.96. Intraobserver coefficient of variation ranged from 0.44 to 2.75. In Bland-Altman analysis, interoperator mean difference ranged from 0.22 to -9.69. An analysis of variance used for repeated measurement was statistically significant for instruments (P < .001) and operators (P = .04), but not for instruments x operators (P = .32).

CONCLUSIONS

The 7 OCT devices presented differing macular thickness measurement values; the lowest value was with Copernicus, and the highest was with Spectralis HRA+OCT (Heidelberg Engineering). The new generation of SD OCT devices has good intraoperator reproducibility, but the Spectralis presents the highest reproducibility together with the best interoperator agreement. The software, and in particular the algorithm, used is the most important factor regarding reproducibility differences in macular thickness measurement. The main result of our study is that macular thickness absolute value differs for each device. For this reason, the devices are not interchangeable.

摘要

目的

评估 6 种新型谱域(SD)光学相干断层扫描(OCT)设备和 1 种时域 OCT 设备的黄斑厚度测量重复性。

设计

前瞻性、观察性研究。

方法

设置:临床实践。研究人群:18 名健康志愿者的 18 只随机选择的连续眼,由 2 名掩蔽操作员使用 6 种 SD-OCT 设备和 1 种时域 OCT 设备评估黄斑厚度。主要观察指标:OCT 测量的中央黄斑厚度的操作者内和操作者间重复性。

结果

黄斑平均厚度范围为 172.77μm 至 272.87μm。观察者内的 Intraclass 系数相关性范围为 0.75 至 0.96。观察者内的变异系数范围为 0.44 至 2.75。在 Bland-Altman 分析中,操作者间的平均差异范围为 0.22μm 至-9.69μm。用于重复测量的方差分析在仪器(P<0.001)和操作者(P=0.04)方面具有统计学意义,但在仪器×操作者(P=0.32)方面无统计学意义。

结论

7 种 OCT 设备呈现出不同的黄斑厚度测量值;Copernicus 的值最低,Spectralis HRA+OCT(海德堡工程)的值最高。新一代 SD-OCT 设备具有良好的操作者内重复性,但 Spectralis 具有最高的重复性和最佳的操作者间一致性。软件,特别是算法,是影响黄斑厚度测量重复性差异的最重要因素。我们研究的主要结果是,每个设备的黄斑厚度绝对值不同。因此,这些设备不可互换。