Srinivasan R, Singh Megha
Indian Institute of Technology-Madras, Tamilnadu 600036, India.
IEEE Trans Biomed Eng. 2004 Oct;51(10):1830-7. doi: 10.1109/TBME.2004.831527.
A laser transillumination tomographic system, consisting of electrical, optical, mechanical, and software components, to obtain multislice images of tissue-equivalent breast phantoms and biological tissues, is developed. The tissue-equivalent phantoms are prepared from paraffin wax mixed with wax color pigments by matching their surface backscattered profiles as measured by multiprobe laser reflectometer, with that of respective tissues. The optical parameters of these phantoms are determined by matching their reflectance profiles with that as obtained by Monte Carlo simulation of optical scattering. For multislice tomographic analysis conical breast phantoms of height 80.0 mm and 80.0 mm base diameter with inclusions of different optical properties and dimensions are developed. The resolution of the inclusions in the tomograms depends on their sizes and optical parameters. The minimum size of the inclusion as detected by this procedure in a slice of diameter 50.0 mm is 3.0 mm. The structural variation as observed in the tomograms of phantoms of combination of biological tissues indicates its possible applications in detecting the abnormalities developing in human healthy soft tissues.
开发了一种激光透照断层扫描系统,该系统由电气、光学、机械和软件组件组成,用于获取组织等效乳腺模型和生物组织的多层图像。组织等效模型是由石蜡与蜡色颜料混合制成的,通过多探头激光反射仪测量其表面后向散射轮廓,并与相应组织的轮廓进行匹配。这些模型的光学参数是通过将其反射率轮廓与光学散射蒙特卡罗模拟得到的轮廓进行匹配来确定的。为了进行多层断层分析,开发了高度为80.0毫米、底部直径为80.0毫米、包含不同光学性质和尺寸内含物的锥形乳腺模型。断层图像中内含物的分辨率取决于它们的大小和光学参数。通过该程序在直径为50.0毫米的切片中检测到的内含物的最小尺寸为3.0毫米。在生物组织组合模型的断层图像中观察到的结构变化表明其在检测人类健康软组织中出现的异常情况方面具有潜在应用。