van Straten Marcel, Venema Henk W, Hartman Joris, den Heeten Gerard J, Grimbergen Cornelis A
Department of Medical Physics, Academic Medical Center, Amsterdam, The Netherlands.
Med Phys. 2004 Oct;31(10):2785-6. doi: 10.1118/1.1796131.
In multi-slice spiral computed tomography (CT) images interpolation artifacts are present. The relationship between the x-ray tube rotation angle and these artifacts is demonstrated. A head phantom was repeatedly scanned with a four-slice CT scanner at different pitch values. Two scans, made with identical scan parameters, nearly always have different x-ray tube starting angles. Consequently, artifacts appeared differently and residual artifacts appeared when subtractions were made. We conclude that only if the x-ray tube starting angle is equal for both scans or if a very low pitch is used, images are highly reproducible.
在多层螺旋计算机断层扫描(CT)图像中存在插值伪影。本文阐述了X射线管旋转角度与这些伪影之间的关系。使用四排CT扫描仪,以不同的螺距值对头部模型进行重复扫描。由于两次扫描的扫描参数相同,但X射线管的起始角度几乎总是不同,因此伪影呈现出不同的形态,在进行相减操作时会出现残余伪影。我们得出的结论是,只有当两次扫描的X射线管起始角度相等或使用非常低的螺距时,图像才具有高度的可重复性。