Dose reduction in computed tomography: the effect of eye and testicle shielding on radiation dose measured in patients with beryllium oxide-based optically stimulated luminescence dosimetry.

作者信息

Grobe Henrik, Sommer Marian, Koch Arne, Hietschold Volker, Henniger Jürgen, Abolmaali Nasreddin

机构信息

OncoRay-Center for Radiation Research in Oncology, Molecular Imaging, Medical Faculty Carl Gustav Carus, Dresden University of Technology, Fetscherstrasse 74, PO Box 86, 01307 Dresden, Germany.

出版信息

Eur Radiol. 2009 May;19(5):1156-60. doi: 10.1007/s00330-008-1241-1. Epub 2008 Dec 11.

Abstract

The aim of this study was to assess the effect of eye and testicle shielding on radiation dose to the lens and the testes of patients undergoing CT examinations. Fifty-one male patients underwent CT twice with identical protocols initially without, the second time with protective garments. Doses to the testes and the lenses were recorded with beryllium oxide-based dosimeters. The dose to the testes and lenses from CT exposure was reduced by 96.2% +/- 1.7% and 28.2% +/- 18.5%, when testicle and eye shielding was used, respectively. The effect of the eye shielding on the eye lens dose was found to depend on the x-ray tube position when the eye is primarily exposed during the scan. The maximum eye lens dose reduction achieved was found to be 43.2% +/- 6.5% corresponding to the anterior position of the tube. A significant correlation between the patient's body mass index and dose exposure could not be found. Eye and testicle shields, apart from being inexpensive and easy to use, were proven to be effective in reducing eye lens and testicle radiation dose burden from CT exposures.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验