Kettenbach Joachim, Helbich Thomas H, Huber Sabine, Zuna Ivan, Dock Wolfgang
Department of Radiology, University Hospital of Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria.
Eur J Radiol. 2005 Feb;53(2):238-44. doi: 10.1016/j.ejrad.2004.04.017.
To objectively quantify the effects of a microbubble contrast agent to differentiate breast tumors with power doppler ultrasound and to compare these results with color doppler ultrasound (CD US).
In 47 patients a microbubble contrast agent was injected intravenously. Computer-assisted quantitative assessment of the color pixel density was performed to evaluate the increase in Doppler signals. Results were compared to previously published results of a color Doppler ultrasound study.
Peak color pixel density at contrast-enhanced power Doppler ultrasound was higher for carcinomas than for benign tumors (P < 0.03). Time to peak enhancement was shorter in carcinomas than in benign tumors (P < 0.01). For both parameters, diagnostic accuracy of power Doppler ultrasound was 69 and 78%, and for color Doppler ultrasound 62 and 76%, respectively.
Quantitative assessment of contrast-enhanced power Doppler ultrasound showed significant differences in malignant and benign breast tumors. Diagnostic accuracy of contrast-enhanced power Doppler ultrasound was higher compared to color Doppler ultrasound.