Quantitative transcutaneous arterial velocity measurements with Doppler flowmeters.

作者信息

Shoor P M, Fronek A, Bernstein E F

出版信息

Arch Surg. 1979 Aug;114(8):922-8. doi: 10.1001/archsurg.1979.01370320054009.

Abstract

A controlled in vitro and in vivo comparative evaluation of four Doppler ultrasonic blood velocimeters was performed to evaluate their ability to measure fluid velocity throughout a wide range of flow rates and velocity profiles, during both steady and pulsatile flow. The instruments tested were the Versatone Bidirectional Doppler M-9, the Sonicaid BV380, the Parks 806 B, and the Delalande DUD 400. All instruments responded linearly to velocities of from 5 to 70 cm/s, and were accurate to within 5% through a wide range of in vitro changes in velocity, pulse frequency, and cross-sectional velocity profile. Quantitative velocity measurements and their derivatives are an inexpensive, simple, and useful tool in the evaluation conditions of patients with peripheral vascular disease, and widespread clinical trials of their value are appropriate at this time.

摘要

对四种多普勒超声血流速度计进行了体外和体内对照比较评估,以评估它们在稳定流和脉动流期间,在很宽的流速和速度分布范围内测量流体速度的能力。所测试的仪器有Versatone双向多普勒M-9、Sonicaid BV380、Parks 806 B和Delalande DUD 400。所有仪器对5至70厘米/秒的速度呈线性响应,并且在体外速度、脉冲频率和横截面速度分布发生广泛变化的情况下,测量精度在5%以内。定量速度测量及其导数在评估外周血管疾病患者的病情时是一种廉价、简单且有用的工具,目前对其价值进行广泛的临床试验是合适的。