Statistical correlation between transient pressure drop and cavitation at closure of a mechanical heart valve.

作者信息

Wu Changfu, Liu Jia-Shing, Hwang Ned H C, Lin Yu-Kweng M

机构信息

Center for Applied Stochastics Research, College of Engineering, Florida Atlantic University, 777 Glades Rd., Boca Raton, FL 33431, USA.

出版信息

ASAIO J. 2005 Jan-Feb;51(1):11-6. doi: 10.1097/01.mat.0000150330.35969.7e.

Abstract

Cavitation on a mechanical heart valve (MHV) is attributable to transient regional pressure drop at the instant of valve closure. As a cavitation bubble collapses, it emits shock waves, which have the characteristics of high frequency oscillations (HFO) on a pressure time trace. The potential for such HFO bursts to cause material damage on an MHV can be measured by the cavitation impulse I, which is defined as the area under the trace of the HFO bursts. In the present study, experiments were conducted on a bileaflet MHV in a durability tester, operated at pulse rates from 300-1,000 bpm. In each case, the transient pressure near an occluder was monitored for 60,000 beats via a transducer. The peak pressure drop Pm and the corresponding cavitation impulse I obtained for the 60,000 beat sequence are found to resemble sample records of two stationary stochastic processes, each of which follows a log normal distribution. Their first order probability density functions are estimated from the records. The correlation is investigated between I and Pm associated with each beat, which is found to be of statistical significance.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验