Bayraktar S, Bayraktar Z
Istanbul Surgery Hospital, Istanbul, Turkey.
Eur J Ophthalmol. 2005 Jan-Feb;15(1):81-8.
To investigate central corneal thickness (CCT) and intraocular pressure (IOP) relationship in eyes with and without previous corneal laser refractive surgery and to compare the estimates of two different tonometers-Goldmann applanation and pneumatonometer.
The study population included 234 glaucoma suspects who were referred to the glaucoma clinic with cup/disc ratios greater than 0.4, asymmetric cupping, and/or IOP greater than 22 mmHg during routine eye examination. Of those, 84 had previous myopic laser-assisted in situ keratomileusis (LASIK) (Group 1) while 150 of them did not (Group 2). CCT was measured by using optical coherence tomography (OCT) and IOP with both Goldmann applanation tonometer (GAT) and pneumatonometer (PT). In both groups, the difference between IOP estimates of two different tonometers and the relationship between CCT and IOP were analyzed.
In eyes with previous LASIK, GAT measured IOP significantly lower than PT (mean difference of 3.8+/-1.9 mmHg, p<0.0001). In eyes with virgin corneas, IOP estimates of GAT or PT were not different from each other (19.9+/-2.8 versus 19.9+/-2.2 mmHg, respectively, p=0.81). In both groups, there was a significant positive correlation between CCT and IOP estimates of GAT(R=0.29, p=0.007 in eyes with LASIK and R=0.38, p<0.0001 in those without), while no similar relationship was present between CCT with those of PT (R=0.03, p=0.76 in eyes with LASIK and R=0.03, p=0.69 in those without).
In eyes with previous LASIK, GAT measured IOP significantly lower than PT. Because IOP estimates of PT were found to be independent from CCT in all of the study eyes, this device was considered to be a more reliable method of IOP estimation than GAT in eyes with and without previous LASIK.
研究有或无角膜激光屈光手术史的眼睛中中央角膜厚度(CCT)与眼压(IOP)的关系,并比较两种不同眼压计——Goldmann压平眼压计和气动眼压计的测量结果。
研究对象包括234名青光眼疑似患者,这些患者在常规眼部检查中因杯盘比大于0.4、杯盘不对称和/或眼压高于22 mmHg而被转诊至青光眼门诊。其中,84人曾接受过近视激光原位角膜磨镶术(LASIK)(第1组),150人未接受过(第2组)。使用光学相干断层扫描(OCT)测量CCT,同时使用Goldmann压平眼压计(GAT)和气动眼压计(PT)测量IOP。在两组中,分析两种不同眼压计测量的IOP差异以及CCT与IOP的关系。
在有LASIK手术史的眼睛中,GAT测量的IOP显著低于PT(平均差异为3.8±1.9 mmHg,p<0.0001)。在角膜未手术的眼睛中,GAT或PT测量的IOP彼此无差异(分别为19.9±2.8 mmHg和19.9±2.2 mmHg,p=0.81)。在两组中,CCT与GAT测量的IOP之间均存在显著正相关(有LASIK手术史的眼睛中R=0.29,p=0.007;无LASIK手术史的眼睛中R=0.38,p<0.0001),而CCT与PT测量的IOP之间无类似关系(有LASIK手术史的眼睛中R=0.03,p=0.76;无LASIK手术史的眼睛中R=0.03,p=0.69)。
在有LASIK手术史的眼睛中,GAT测量的IOP显著低于PT。由于在所有研究眼中发现PT测量的IOP与CCT无关,因此在有或无LASIK手术史的眼睛中,该设备被认为是比GAT更可靠的IOP测量方法。