Krueger Sascha, Timinger Holger, Grewer Ruediger, Borgert Joern
Philips Research Laboratories, Division Technical Systems, Roentgenstrasse 24-26, 22335 Hamburg, Germany.
Phys Med Biol. 2005 Feb 21;50(4):581-97. doi: 10.1088/0031-9155/50/4/002.
A novel setup for the integration of a magnetic tracking system (MTS) into a catheter laboratory environment by attaching the field generator of the MTS to the C-arm of the x-ray system was investigated. The metal interference of the x-ray system with the MTS then becomes static and can be calibrated. The registration of a tracked catheter to 2D live x-ray images with high and well-defined accuracy is thus possible for arbitrary C-arm positions. Interference with imaging by blocking the field of view is avoided. Real-time registration methods to maintain the registration of the catheter also to static 2D or 3D images (roadmaps), regardless of the C-arm rotation during catheter tracking, were investigated. Residual registration errors of the tracked catheter with respect to the static roadmaps (2D and 3D) were reduced by using the vessel geometry and shape information. The algorithms potentially allow for motion correction (e.g. due to respiration). Using the shape-based correction algorithms the average registration accuracies to static roadmaps for different C-arm angulations and catheter positions were determined to be 3.3-4.2 mm. The magnetic registration of the C-arm was furthermore allowed to compute the imaging geometry (position of the C-arm) and to produce virtual angiographic preview images before contrast injection and x-ray dose application. Thus, optimal projection geometries and collimator settings for the target region can be chosen in an 'off line' fashion. The proposed MTS-supported navigation setup on both 3D (static) and 2D (live and static) roadmaps merges the high resolution and real-time feedback of 2D x-ray images with the navigation support from 3D static images.
研究了一种通过将磁跟踪系统(MTS)的场发生器连接到X射线系统的C形臂,从而将磁跟踪系统集成到导管实验室环境中的新型设置。X射线系统对MTS的金属干扰随后变为静态,可以进行校准。因此,对于任意C形臂位置,都可以将跟踪的导管以高且明确的精度配准到二维实时X射线图像。避免了因遮挡视野而对成像造成的干扰。还研究了实时配准方法,以在导管跟踪过程中无论C形臂如何旋转,都能将导管与静态二维或三维图像(路线图)保持配准。通过使用血管几何形状和形状信息,减少了跟踪导管相对于静态路线图(二维和三维)的残余配准误差。这些算法可能允许进行运动校正(例如由于呼吸)。使用基于形状的校正算法,确定了不同C形臂角度和导管位置下到静态路线图的平均配准精度为3.3 - 4.2毫米。此外,C形臂的磁配准还允许计算成像几何形状(C形臂的位置),并在注入造影剂和施加X射线剂量之前生成虚拟血管造影预览图像。因此,可以以“离线”方式选择目标区域的最佳投影几何形状和准直器设置。所提出的在三维(静态)和二维(实时和静态)路线图上均由MTS支持的导航设置,将二维X射线图像的高分辨率和实时反馈与三维静态图像的导航支持相结合。