Pathak Chetna, Van Horn Mark, Weeks Susan, Bullitt Elizabeth
Department of Surgery, University of North Carolina, Chapel Hill, NC, USA.
Med Image Comput Comput Assist Interv. 2005;8(Pt 2):239-46. doi: 10.1007/11566489_30.
Accurate 3D/2D vessel registration is complicated by issues of image quality, occlusion, and other problems. This study performs a quantitative comparison of 3D/2D vessel registration in which vessels segmented from preoperative CT or MR are registered with biplane x-ray angiograms by either a) simultaneous two-view registration with advance calculation of the relative pose of the two views, or b) sequential registration with each view. We conclude on the basis of phantom studies that, even in the absence of image errors, simultaneous two-view registration is more accurate than sequential registration. In more complex settings, including clinical conditions, the relative accuracy of simultaneous two-view registration is even greater.
准确的三维/二维血管配准因图像质量、遮挡及其他问题而变得复杂。本研究对三维/二维血管配准进行了定量比较,其中从术前CT或MR分割出的血管通过以下两种方式与双平面X射线血管造影进行配准:a)通过预先计算两个视图的相对位姿进行同步双视图配准,或b)对每个视图进行顺序配准。我们基于模型研究得出结论,即使在没有图像误差的情况下,同步双视图配准也比顺序配准更准确。在包括临床情况在内的更复杂场景中,同步双视图配准的相对准确性更高。