Tampoia Marilina, Brescia Vincenzo, Fontana Antonietta, Maggiolini Piera, Lapadula Giovanni, Pansini Nicola
Unità Operativa di Patologia Clinica I, Policlinico di Bari, Piazza G. Cesare 11, 70100 Bari, Italy.
Clin Chim Acta. 2005 May;355(1-2):137-44. doi: 10.1016/j.cccn.2004.12.017.
Autoantibodies against cyclic citrullinated peptide (anti-CCP) are considered to be a sensitive and specific marker for rheumatoid arthritis (RA). This study evaluated the analytical performance and clinical correlation of an automated enzyme immunoassay (DSX, DINEX Technologies), for the detection of anti-CCP autoantibodies (DIASTAT anti-CCP, Axis-Shield, DUNDEE UK).
Commercial controls and serum pools were used to determine its precision, analytical sensitivity, functional sensitivity and linearity. Sera from 83 patients with established RA and from 140 controls, including patients with various autoimmune diseases, viral infections and cancer, as well as sex- and age-matched healthy subjects, were studied. The rheumatoid factor (RF) was also assayed in each sample, and the results were compared to the anti-CCP findings.
The total imprecision (CV%) was 4.7-7.2% for concentrations ranging between 1.98 and 71.81 U/mL. The lower detection limit was 0.038 U/mL. At a cut-off of 5 U/mL, the sensitivity and specificity for RA were 67.5% and 99.3%, respectively. The RF had a sensitivity of 66.3% and a lower specificity 82.1% than anti-CCP. When the two antibodies were used together, the specificity was 99.1%.
The anti-CCP assay we examined on a fully automated system showed a good analytical performance (analytical and functional sensitivity, linearity) and good clinical correlation. We conclude that this system can provide rapid, useful data.
抗环瓜氨酸肽自身抗体(抗CCP)被认为是类风湿关节炎(RA)的一种敏感且特异的标志物。本研究评估了一种用于检测抗CCP自身抗体(DIASTAT抗CCP,Axis-Shield,英国邓迪)的自动化酶免疫测定法(DSX,DINEX Technologies)的分析性能及临床相关性。
使用商业对照品和血清池来确定其精密度、分析灵敏度、功能灵敏度和线性。研究了83例确诊RA患者以及140例对照者的血清,对照者包括患有各种自身免疫性疾病、病毒感染和癌症的患者以及年龄和性别匹配的健康受试者。对每个样本也检测了类风湿因子(RF),并将结果与抗CCP检测结果进行比较。
浓度在1.98至71.81 U/mL之间时,总不精密度(CV%)为4.7 - 7.2%。检测下限为0.038 U/mL。在5 U/mL的临界值时,RA的灵敏度和特异性分别为67.5%和99.3%。RF的灵敏度为66.3%且特异性低于抗CCP,为82.1%。当两种抗体联合使用时,特异性为99.1%。
我们在全自动系统上检测的抗CCP检测方法显示出良好的分析性能(分析和功能灵敏度、线性)以及良好的临床相关性。我们得出结论,该系统能够提供快速、有用的数据。