Djajaputra David, Nehru Ramasamy, Bruch Philip M, Ayyangar Komanduri M, Raman Natarajan V, Enke Charles A
Department of Radiation Oncology, University of Nebraska Medical Center, 987521 Nebraska Medical Center, Omaha, NE 68198-7521, USA.
Phys Med Biol. 2005 May 7;50(9):N93-9. doi: 10.1088/0031-9155/50/9/N03. Epub 2005 Apr 13.
Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag.