Arditi Aries
Arlene R. Gordon Research Institute, Lighthouse International, New York, NY 10022, USA.
Invest Ophthalmol Vis Sci. 2005 Jun;46(6):2225-9. doi: 10.1167/iovs.04-1198.
To develop a small-format letter contrast sensitivity test with improved accuracy, suitable for portable use and near testing, and having a simple and unambiguous scoring system that can be used with reference to existing norms.
A near test, printed on resin-coated paper and mounted on plastic was developed by using Sloan letters, with 0.04 log unit contrast decrements between each letter and a simple scoring procedure. Monte Carlo methods and a Weibull function model of visual performance were used to assess test accuracy.
The new test has 28% lower score standard deviations than the Pelli-Robson Contrast Sensitivity Chart, over a wide range of low misreporting rates, while maintaining test scores that differ from the Pelli-Robson by 0.01 log unit, permitting use of norms collected with the Pelli-Robson test.
While maintaining comparability with Pelli-Robson norms, the new test has improved accuracy in comparison with the Pelli-Robson chart and several other advantages that result from its smaller size.
开发一种小型字母对比敏感度测试,提高准确性,适合便携式使用和近视力测试,并具有简单明确的评分系统,可参照现有标准使用。
采用斯隆字母,在树脂涂层纸上打印并安装在塑料上制作近视力测试,每个字母之间的对比度递减0.04对数单位,并采用简单的评分程序。使用蒙特卡罗方法和视觉性能的威布尔函数模型评估测试准确性。
在广泛的低误报率范围内,新测试的分数标准差比斯隆对比敏感度图表低28%,同时保持与斯隆测试相差0.01对数单位的测试分数,允许使用通过斯隆测试收集的标准。
在保持与斯隆标准可比性的同时,新测试与斯隆图表相比提高了准确性,并且因其较小尺寸还具有其他几个优点。