Suppr超能文献

改进字母对比敏感度测试的设计。

Improving the design of the letter contrast sensitivity test.

作者信息

Arditi Aries

机构信息

Arlene R. Gordon Research Institute, Lighthouse International, New York, NY 10022, USA.

出版信息

Invest Ophthalmol Vis Sci. 2005 Jun;46(6):2225-9. doi: 10.1167/iovs.04-1198.

Abstract

PURPOSE

To develop a small-format letter contrast sensitivity test with improved accuracy, suitable for portable use and near testing, and having a simple and unambiguous scoring system that can be used with reference to existing norms.

METHODS

A near test, printed on resin-coated paper and mounted on plastic was developed by using Sloan letters, with 0.04 log unit contrast decrements between each letter and a simple scoring procedure. Monte Carlo methods and a Weibull function model of visual performance were used to assess test accuracy.

RESULTS

The new test has 28% lower score standard deviations than the Pelli-Robson Contrast Sensitivity Chart, over a wide range of low misreporting rates, while maintaining test scores that differ from the Pelli-Robson by 0.01 log unit, permitting use of norms collected with the Pelli-Robson test.

CONCLUSIONS

While maintaining comparability with Pelli-Robson norms, the new test has improved accuracy in comparison with the Pelli-Robson chart and several other advantages that result from its smaller size.

摘要

目的

开发一种小型字母对比敏感度测试,提高准确性,适合便携式使用和近视力测试,并具有简单明确的评分系统,可参照现有标准使用。

方法

采用斯隆字母,在树脂涂层纸上打印并安装在塑料上制作近视力测试,每个字母之间的对比度递减0.04对数单位,并采用简单的评分程序。使用蒙特卡罗方法和视觉性能的威布尔函数模型评估测试准确性。

结果

在广泛的低误报率范围内,新测试的分数标准差比斯隆对比敏感度图表低28%,同时保持与斯隆测试相差0.01对数单位的测试分数,允许使用通过斯隆测试收集的标准。

结论

在保持与斯隆标准可比性的同时,新测试与斯隆图表相比提高了准确性,并且因其较小尺寸还具有其他几个优点。