Bogucarskis K, Salmins A, Gfirtner H, Anatschkowa E
Radiation Safety Centre, Riga, Latvia.
Radiat Prot Dosimetry. 2005;114(1-3):176-9. doi: 10.1093/rpd/nch553.
This article deals with the estimation of doses received by patients undergoing radiological examinations in order to establish dose reference levels (DRLs) in Latvia. Several large hospitals, small hospitals and private practices were selected for patient dose measurements. The measurements were carried out using calibrated thermoluminiscence dosemeters attached to the patient's skin. Exposure parameters and patient's data were recorded. The entrance surface doses (ESDs) to patients undergoing several common X-ray examinations (chest AP/PA, chest LAT, lumbar spine AP/PA, lumbar spine LAT and pelvis) were measured. Data concerning the kV(p) settings, used type of films, focus-film distance and the ESD values were analysed and compared with those recommended by the European Community (EC). Among the different hospitals and private practices, discrepancies in the patient doses and techniques used for the examination were found, where the doses exceeded the EC recommended values owing to a very low kV(p) and a very low sensitivity of the screen film combinations used.
本文旨在通过对接受放射检查的患者所受剂量进行估算,以确定拉脱维亚的剂量参考水平(DRL)。为此,选取了几家大型医院、小型医院和私人诊所进行患者剂量测量。测量采用附着于患者皮肤的校准热释光剂量计进行。同时记录曝光参数和患者数据。对接受几种常见X射线检查(胸部前后位/后前位、胸部侧位、腰椎前后位/后前位、腰椎侧位和骨盆)的患者的体表入射剂量(ESD)进行了测量。分析了有关千伏(峰值)设置、所用胶片类型、焦片距和ESD值的数据,并与欧洲共同体(EC)推荐的值进行了比较。在不同的医院和私人诊所中,发现患者剂量和检查所用技术存在差异,由于千伏(峰值)极低以及所用增感屏-胶片组合的灵敏度极低,剂量超过了EC推荐值。