Bax Liesbeth, Bakker Chris J G, Klein Willemijn M, Blanken Niels, Beutler Jaap J, Mali Willem P T R M
Department of Radiology, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
J Vasc Interv Radiol. 2005 Jun;16(6):807-14. doi: 10.1097/01.RVI.0000161144.98350.28.
To assess the validity and the direct, short-term, and long-term reproducibility of renal blood flow (RBF) measurements with phase-contrast (PC) magnetic resonance (MR) imaging.
In 20 healthy volunteers, RBF measurements were repeated with and without repositioning. Internal validity was assessed by comparing the total RBF with the difference in aortic flow above and below the renal arteries. In 19 healthy volunteers, RBF measurements were performed at two different occasions. In 40 healthy volunteers, RBF measurements were performed to assess normal values as a function of age. Analyses were performed according to Bland and Altman.
The technical success rate ranged from 78% to 85%. Total RBF and the difference in aortic flow rates showed good agreement (Pearson correlation coefficient, 0.72; P = .002). Directly repeated measurements had a mean difference of 54 mL/min in total RBF with a coefficient of variation (CV) of 17%. For repeated measurements with repositioning, the mean difference in total RBF was 74 mL/min (CV, 23%). Repeated measurements on different occasions showed a CV of 20%. The mean total RBF of the 40 healthy volunteers was 838 mL/min +/- 244 (SD).
RBF measurement with PC MR has a success rate greater than 75%. The demonstrated internal reliability of this method and fair reproducibility of the flow parameters is crucial for further studies of the renal artery with MR imaging.
评估利用相位对比(PC)磁共振(MR)成像测量肾血流量(RBF)的有效性以及直接、短期和长期的可重复性。
对20名健康志愿者进行了有无重新定位情况下的RBF测量重复操作。通过比较总RBF与肾动脉上下主动脉血流量差值来评估内部有效性。对19名健康志愿者在两个不同时间点进行了RBF测量。对40名健康志愿者进行RBF测量以评估其作为年龄函数的正常值。根据布兰德和奥特曼方法进行分析。
技术成功率在78%至85%之间。总RBF与主动脉血流速率差值显示出良好的一致性(皮尔逊相关系数为0.72;P = 0.002)。直接重复测量时,总RBF的平均差值为54 mL/min,变异系数(CV)为17%。重新定位后重复测量时,总RBF的平均差值为74 mL/min(CV为23%)。在不同时间点重复测量的CV为20%。40名健康志愿者的平均总RBF为838 mL/min ± 244(标准差)。
利用PC MR测量RBF的成功率大于75%。该方法所展示的内部可靠性以及血流参数的尚可重复性对于利用MR成像进一步研究肾动脉至关重要。