Yazici A Rüya, Baseren Meserret, Gokalp Saadet
Department of Conservative Dentistry, Faculty of Dentistry, Hacettepe University, Ankara, Turkey.
Quintessence Int. 2005 Jun;36(6):417-22.
The aim of this in vitro study was to compare the performance of a laser fluorescence device (DIAGNOdent) and a caries-detector dye for detecting residual dentinal caries.
Forty extracted human molar teeth with deep dentinal carious lesions on their approximal surfaces were used for the study. The caries were removed by a new chemomechanical system, Carisolv. Three independent observers checked all cavities for residual caries using caries-detector dye and DIAGNOdent. The presence or absence of residual caries was verified using polarized light microscopy as the gold standard. Three observers recorded two different measurements at the same site after 2 weeks to determine intra- and interexaminer reproducibility of the DIAGNOdent.
Average kappa values of 0.61 for interexaminer reproducibility and 0.94 for intraexaminer reproducibility were determined. The specificity of caries-detector dye was found to be 100% and the specificity of the DIAGNOdent was 86%.
Although DIAGNOdent showed greater sensitivity than caries-detector dye (P < .05), low sensitivity values for each diagnostic system may lead to teeth with residual caries being undetected. Therefore, it might be advisable to reconsider the criteria for detecting residual caries during tooth preparation.
本体外研究旨在比较激光荧光装置(DIAGNOdent)和龋病检测染料在检测残留牙本质龋方面的性能。
本研究使用了40颗在邻面有深牙本质龋损的拔除人类磨牙。龋坏组织通过一种新的化学机械系统Carisolv去除。三名独立观察者使用龋病检测染料和DIAGNOdent检查所有窝洞是否存在残留龋。使用偏光显微镜作为金标准来验证是否存在残留龋。三名观察者在2周后在同一部位记录了两种不同的测量结果,以确定DIAGNOdent的检查者间和检查者内的可重复性。
确定检查者间可重复性的平均kappa值为0.61,检查者内可重复性的平均kappa值为0.94。发现龋病检测染料的特异性为100%,DIAGNOdent的特异性为86%。
尽管DIAGNOdent显示出比龋病检测染料更高的敏感性(P <.05),但每个诊断系统的低敏感性值可能导致残留龋的牙齿未被检测到。因此,在备牙过程中重新考虑检测残留龋的标准可能是明智的。