Armbruster David A, Alexander Debra B
Scientific Affairs, Abbott Laboratories, 100 Abbott Park Road, Abbott Park, IL 60046, USA.
Clin Chim Acta. 2006 Nov;373(1-2):37-43. doi: 10.1016/j.cca.2006.04.022. Epub 2006 Jun 14.
Integrated systems that combine clinical chemistry and immunoassay analyzers are used routinely. Sample to sample carryover is an inherent risk and can cause erroneously high patient test results for immunoassays. IVD manufacturers and laboratories must be aware of this phenomenon and guard against it.
We used a sample carryover protocol that directs the clinical chemistry module to process samples with very high immunoassay analyte concentrations followed by samples with very low concentrations for the same analyte. Low concentration samples were then tested by the immunoassay module to determine if the clinical chemistry module caused primary sample tube to primary sample tube carryover of the immunoassay analyte.
Sample carryover was assessed on the Abbott ci8200 for HBsAg, AFP, beta-hCG, and PSA. Observed HBsAg carryover met the design specification of <0.1 ppm. Carryover for the other analytes was <0.1 ppm or below the assay limit of detection.
IVD manufacturers must design integrated systems to minimize primary specimen tube carryover and avoid analytical laboratory error that can impact patient safety. Carryover testing is difficult for clinical laboratories to perform in order to verify system performance. Laboratories must consider the potential for specimen carryover and its impact on results whether moving primary sample tubes between separate analyzers or using an integrated system.
结合临床化学分析仪和免疫分析仪器的集成系统已被常规使用。样本间携带污染是一种内在风险,可能导致免疫分析中患者检测结果出现错误的高值。体外诊断制造商和实验室必须意识到这种现象并加以防范。
我们采用了一种样本携带污染方案,指导临床化学模块先处理免疫分析物浓度非常高的样本,然后处理相同分析物浓度非常低的样本。接着,免疫分析模块对低浓度样本进行检测,以确定临床化学模块是否导致了免疫分析物在原样本管之间的携带污染。
在雅培ci8200上对乙肝表面抗原(HBsAg)、甲胎蛋白(AFP)、β-人绒毛膜促性腺激素(β-hCG)和前列腺特异性抗原(PSA)进行了样本携带污染评估。观察到的HBsAg携带污染符合<0.1 ppm的设计规格。其他分析物的携带污染<0.1 ppm或低于检测限。
体外诊断制造商必须设计集成系统,以尽量减少原样本管的携带污染,并避免可能影响患者安全的分析实验室误差。临床实验室难以进行携带污染检测以验证系统性能。无论是在不同分析仪之间转移原样本管还是使用集成系统,实验室都必须考虑样本携带污染的可能性及其对结果的影响。