Dietlein Thomas S, Roessler Gernot, Lüke Christoph, Dinslage Sven, Roters Sigrid, Jacobi Philipp C, Walter Peter, Krieglstein Guenter K
Department of Ophthalmology, University of Cologne, D-50931 Cologne, Germany.
J Cataract Refract Surg. 2005 May;31(5):1006-10. doi: 10.1016/j.jcrs.2004.09.049.
To assess the practical feasibility and signal quality of axial length measurements by partial coherence laser interferometry in silicone oil-filled eyes with previous complicated vitreoretinal surgery.
Department of Ophthalmology, University Cologne, Cologne, Germany.
Using a Zeiss IOLMaster, axial length measurements and signal-to-noise ratios of optical biometry in silicone oil-filled eyes (n=45) and contralateral eyes without tamponade (n=41) were analyzed.
Axial length measurements with signal-to-noise ratio > or =2 were feasible in 41 of 45 silicone oil-filled eyes (91%) and 37 of 41 eyes without tamponade (90%). Cataract, central retinal detachment, vitreous hemorrhage, and emulsified oil droplets attached to the intraocular lens were reasons for failure of partial coherence laser interferometry. The signal-to-noise ratio of the first 2 measurements was significantly smaller (P=.04) in silicone-filled eyes (4.4 +/- 2.0) than in eyes without tamponade (5.5 +/- 3.0). Axial lengths of the oil-filled eye and the contralateral eye showed a significant intraindividual correlation (P<.0001, Spearman r=0.84).
Partial coherence laser interferometry shows good clinical practicability in silicone oil-filled eyes with previous complicated vitreoretinal surgery. Further studies are needed to assess the reliability of these measurements with regard to postoperative refraction after combined oil removal and cataract surgery.
评估在曾接受复杂玻璃体视网膜手术的硅油填充眼中,使用部分相干激光干涉测量法测量眼轴长度的实际可行性和信号质量。
德国科隆大学眼科。
使用蔡司IOLMaster,分析硅油填充眼(n = 45)和对侧未行填塞的眼(n = 41)的眼轴长度测量值及光学生物测量的信噪比。
在45只硅油填充眼中,41只(91%)眼轴长度测量的信噪比≥2是可行的;在41只未行填塞的眼中,37只(90%)可行。白内障、视网膜中央脱离、玻璃体积血以及附着于人工晶状体的乳化油滴是部分相干激光干涉测量法失败的原因。硅油填充眼首次测量的前两次测量的信噪比(4.4±2.0)显著低于未行填塞的眼(5.5±3.0)(P = 0.04)。硅油填充眼和对侧眼的眼轴长度显示出显著的个体内相关性(P<0.0001,Spearman相关系数r = 0.84)。
对于曾接受复杂玻璃体视网膜手术的硅油填充眼,部分相干激光干涉测量法具有良好的临床实用性。需要进一步研究以评估在联合硅油取出和白内障手术后这些测量对于术后屈光的可靠性。