Rapid identification of Mycobacterium avium complex (MAC) is possible by use of AccuProbe (Gen-Probe, San Diego, Calif.). To evaluate the reliability of the MAC AccuProbe for testing 7H9 cultures inoculated with broth from MGIT cultures positive for acid-fast bacilli or growth on a solid medium, we compared probe results to results obtained by sequencing a portion of the 16S rRNA gene. Isolates were sequenced if the MAC probe result was <100,000 relative light units (RLU) or if it was > or =100,000 RLU and the colony morphology was not classic or there were two colony types. For the 1,389 cultures tested in phase 1, conducted to evaluate cutoff values for the MAC probe in testing of 7H9 cultures inoculated with broth from MGIT cultures, the sensitivity and specificity of the MAC AccuProbe were 97.7% and 88.8%, respectively, according to the manufacturer's interpretive criteria (> or =30,000 RLU is positive). If the cutoff for a positive result were 80,000 RLU, the specificity would be 100% and the sensitivity 92.3%. Of the 344 isolates in phase 2, which was conducted to confirm the 80,000-RLU cutoff for a positive result and therefore included only isolates with a MAC probe result of < or =100,000 RLU, 13 of 16 with results of > or =30,000 but <80,000 RLU were identified as mycobacteria other than MAC, including five Mycobacterium tuberculosis complex isolates. These data support the use of 80,000 RLU as the cutoff for a positive result in testing of 7H9 broth cultures with the MAC AccuProbe.
使用AccuProbe(基因探针,加利福尼亚州圣地亚哥市)能够快速鉴定鸟分枝杆菌复合群(MAC)。为了评估MAC AccuProbe检测接种自耐酸杆菌MGIT培养肉汤阳性或在固体培养基上生长的7H9培养物的可靠性,我们将探针检测结果与通过对16S rRNA基因的一部分进行测序获得的结果进行了比较。如果MAC探针检测结果<100,000相对光单位(RLU),或者结果≥100,000 RLU且菌落形态不典型或存在两种菌落类型,则对分离株进行测序。在第1阶段测试的1389份培养物中,为评估MAC探针在检测接种自MGIT培养肉汤的7H9培养物时的临界值,根据制造商的解释标准(≥30,000 RLU为阳性),MAC AccuProbe的敏感性和特异性分别为97.7%和88.8%。如果阳性结果的临界值为80,000 RLU,则特异性将为100%,敏感性为92.3%。在第2阶段的344份分离株中,该阶段旨在确认阳性结果的80,000-RLU临界值,因此仅包括MAC探针结果≤100,000 RLU的分离株,16份结果≥30,000但<80,000 RLU的分离株中有13份被鉴定为非MAC分枝杆菌,包括5份结核分枝杆菌复合群分离株。这些数据支持将80,000 RLU用作使用MAC AccuProbe检测7H9肉汤培养物时阳性结果的临界值。