van Weert Stijn, Stokroos Robert J, Rikers Mirçea M J G, van Dijk Pim
Department of Otorhinolaryngology and Head & Neck Surgery, University Hospital Maastricht, Maastricht, The Netherlands.
Acta Otolaryngol. 2005 Jul;125(7):725-31. doi: 10.1080/00016480510028492.
There was no evidence that a reduction in current was needed for nerve stimulation as a result of modiolar placement of a cochlear implant. However, modiolus hugging did reduce the spread of excitation for the basal and apical electrodes. This improved stimulation selectivity may result in improved speech discrimination by implant users.
To test the effect of modiolar placement of a cochlear implant on stimulation thresholds, and to confirm whether peri-modiolar electrode placement resulted in the hypothesized reduced spread of excitation.
Auditory nerve responses were measured by means of neural response telemetry (NRT) in 14 subjects. All subjects received a Nucleus CI24R(CS) Contour implant. For each subject, the stimulation threshold and response growth rate were determined on all the odd-numbered electrodes, using a masker-and-probe paradigm. In addition, the spatial spread of excitation was measured on electrodes 1, 6, 11, 16 and 20, using a variable-masker paradigm. All NRT measurements were performed intra-operatively, both before and after peri-modular placement of the electrode by removal of surgical stylet.
Removal of the stylet had no significant effect on the threshold and growth rate of NRT responses. It caused a reduction in the spread of excitation for electrodes 1, 6 and 20, but not for electrodes 11 and 16.
没有证据表明由于人工耳蜗蜗轴植入,神经刺激时需要降低电流。然而,蜗轴贴合确实减少了基底电极和顶电极的兴奋扩散。这种改善的刺激选择性可能会提高植入者的言语辨别能力。
测试人工耳蜗蜗轴植入对刺激阈值的影响,并确认蜗轴周围电极植入是否会导致假设的兴奋扩散减少。
通过神经反应遥测(NRT)测量14名受试者的听神经反应。所有受试者均接受了Nucleus CI24R(CS) Contour植入物。对于每名受试者,使用掩蔽器-探针范式在所有奇数电极上确定刺激阈值和反应增长率。此外,使用可变掩蔽器范式在电极1、6、11、16和20上测量兴奋的空间扩散。所有NRT测量均在手术中通过移除手术探针在电极蜗轴周围植入前后进行。
移除探针对NRT反应的阈值和增长率没有显著影响。它导致电极1、6和20的兴奋扩散减少,但电极11和16没有。