Bohannon Richard W
Department of Physical Therapy, School of Allied Health, U-2101, University of Connecticut, Storrs, CT 06269-2101, USA.
Percept Mot Skills. 2005 Jun;100(3 Pt 1):795-8. doi: 10.2466/pms.100.3.795-798.
Repeated measures of grip strength obtained bilaterally with a Jamar and a MicroFET 4 dynamometer were compared. Measurements obtained with the MicroFET 4 tended to be slightly (2.2-3.1 lb.) higher but were highly correlated (r > or = .96) with those obtained with the Jamar. Parallel reliability for the two devices was excellent (intraclass correlation coefficient > or = .96). Although clinicians should be cautious about using the devices interchangeably, the MicroFET 4 appears to be a legitimate alternative to the Jamar dynamometer.