Maia Ana F, Caldas Linda V E
Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, Av. Prof. Lineu Prestes, 2242, CEP 05508-000, São Paulo, Brazil.
Phys Med Biol. 2005 Aug 21;50(16):3837-47. doi: 10.1088/0031-9155/50/16/012. Epub 2005 Aug 2.
A special parallel-plate ionization chamber was developed. The motivation for the construction of this new chamber was mainly to fulfil the need of a reference system for computed tomography standard beams in the Calibration Laboratory of IPEN. However, the chamber was tested also in standard radiation beams of mammography and conventional diagnostic radiology. The chamber was manufactured at the institute workshop, as simply and cheaply as possible. Its design differs from the common ionization chambers used in dosimetric procedures of computed tomography equipment, because it is a parallel-plate chamber instead of a cylindrical chamber. However, its dimensions and sensitive volume are very similar to those of a commercial pencil ionization chamber. The new ionization chamber was submitted to several characterization and quality control tests, showing its very good performance.