False-positive alarms for bacterial screening of platelet concentrates with BacT/ALERT new-generation plastic bottles: a multicenter pilot study.

作者信息

Hundhausen T, Müller T H

机构信息

DRK-Blutspendedienst NSTOB, Institute Springe, Germany.

出版信息

Transfusion. 2005 Aug;45(8):1267-74. doi: 10.1111/j.1537-2995.2005.00194.x.

Abstract

BACKGROUND

The microbial detection system BacT/ALERT (bioMérieux) is widely used to monitor bacterial contamination of platelet concentrates (PCs). Recently, the manufacturer introduced polycarbonate culture bottles and a modified pH-sensitive liquid emulsion sensor as microbial growth indicator. This reconfigured assay was investigated in a routine setting.

STUDY DESIGN AND METHODS

In each of eight transfusion centers, samples from 500 consecutive PCs were monitored for 1 week. For all PCs with a positive BacT/ALERT signal, retained samples and, if available, original PC containers and concomitant red blood cell concentrates were analyzed independently. Initially BacT/ALERT-positive PCs without bacterial identification in any sample were defined as false-positive. BacT/ALERT-positive PCs with bacteria in the first sample only were called potentially positive. PCs with bacteria in the first sample and the same strain in at least one additional sample were accepted as positive.

RESULTS

Five PCs (0.13%) were positive, 9 PCs (0.23%) were potentially positive, and 35 PCs (0.9%) were false-positive. The rate of false-positive BacT/ALERT results varied substantially between centers (<0.2%-3.2%). Tracings from false-positive cultures lacked an exponential increase of the signal during incubation. Most of these false-positives were due to malfunctioning cells in various BacT/ALERT incubation units.

CONCLUSION

Careful assessment of individual tracings of samples with positive signals helps to identify malfunctioning incubation units. Their early shutdown or replacement minimizes the high rate of unrectifiable product rejects attributed to false-positive alarms and avoids unnecessary concern of doctors and patients after conversion to a reconfigured BacT/ALERT assay.

摘要

背景

微生物检测系统BacT/ALERT(生物梅里埃公司)被广泛用于监测血小板浓缩物(PC)的细菌污染情况。最近,制造商推出了聚碳酸酯培养瓶和一种经过改良的对pH敏感的液体乳液传感器作为微生物生长指示剂。在常规环境中对这种重新配置的检测方法进行了研究。

研究设计与方法

在八个输血中心中的每一个,对500份连续的PC样本进行了为期1周的监测。对于所有BacT/ALERT信号呈阳性的PC,对留存样本以及(如有)原始PC容器和同时期的红细胞浓缩物进行独立分析。最初,在任何样本中均未鉴定出细菌的BacT/ALERT阳性PC被定义为假阳性。仅在首个样本中检测到细菌的BacT/ALERT阳性PC被称为潜在阳性。首个样本中检测到细菌且至少在一个其他样本中检测到相同菌株的PC被判定为阳性。

结果

5份PC(0.13%)呈阳性,9份PC(0.23%)为潜在阳性,35份PC(0.9%)为假阳性。BacT/ALERT假阳性结果的发生率在各中心之间差异很大(<0.2% - 3.2%)。假阳性培养物的曲线在孵育期间信号没有呈指数增加。这些假阳性大多是由于各种BacT/ALERT孵育单元中的细胞故障所致。

结论

仔细评估信号呈阳性的样本的个体曲线有助于识别故障孵育单元。尽早关闭或更换这些单元可将因假阳性警报导致的无法纠正的产品拒收率降至最低,并避免在转换为重新配置的BacT/ALERT检测方法后医生和患者的不必要担忧。