Augousti A T, Maletras F-X, Mason J
School of Life Sciences, Kingston University, Surrey, UK.
Physiol Meas. 2005 Oct;26(5):585-90. doi: 10.1088/0967-3334/26/5/001. Epub 2005 May 31.
The macro-bending loss effect in optical fibres has been used to redevelop a sensor for the measurement of thoracic and abdominal circumferences in non-invasive respiratory monitoring. The new sensor uses a novel figure-of-eight loop configuration, which results in increased linearity of response, less mechanical resistance and hysteresis, as well as other benefits. The performance of the new sensor as applied to respiratory monitoring is examined, and indicates a higher resolution and sensitivity than the old. This enhanced performance enables measurement of respiratory and cardiac function using the same transducing fibre.
光纤中的宏弯曲损耗效应已被用于重新开发一种传感器,用于在无创呼吸监测中测量胸腹部周长。这种新型传感器采用了一种新颖的八字形环配置,这使得响应线性度提高、机械阻力和滞后减小,以及带来其他好处。研究了这种新型传感器应用于呼吸监测时的性能,结果表明其分辨率和灵敏度比旧传感器更高。这种增强的性能使得能够使用同一传感光纤测量呼吸和心脏功能。