To establish the interobserver reliability for measuring neck mobility in human subjects by means of an electromagnetic tracking device, the Flock-of-Birds system.
Two observers independently and in random order assessed the cervical range-of-motion in 30 subjects with a dysfunction in the neck and shoulder region (symptomatic subjects) and 30 subjects without known pathology (asymptomatic subjects). Measurements included rotation in neutral position, in flexed position and in extended position, flexion-extension, and lateral bending (all active and passive). Reliability was analyzed by intraclass correlation coefficient (ICC) and agreement by limits of agreement and percentage of paired observations within 5 degrees, 10 degrees, and 15 degrees.
For asymptomatic subjects, the ICC varied from 0.57 to 0.85, and the limits of agreement varied between 14.5 degrees and 27.0 degrees. The percentage of paired observations within 5 degrees, 10 degrees, and 15 degrees ranged from 31% to 57%, 58% to 90%, and 78% to 93%, respectively. For symptomatic subjects, the ICC varied from 0.36 to 0.91, and the limits of agreement varied between 9.6 degrees and 37.8 degrees. The percentage of paired observations within 5 degrees, 10 degrees, and 15 degrees ranged from 17% to 60%, 33% to 93%, and 50% to 97%, respectively.
Despite the use of a standardized protocol and a sophisticated measurement system, the interobserver reliability of neck mobility was variable in quality, with reliability being good in rotation in neutral position, flexion-extension, and lateral bending.
通过电磁跟踪设备“鸟类群系统”建立人体颈部活动度测量的观察者间可靠性。
两名观察者独立且随机地评估了30名颈部和肩部区域功能障碍的受试者(有症状受试者)以及30名无已知病变的受试者(无症状受试者)的颈椎活动范围。测量包括中立位、屈曲位和伸展位的旋转、屈伸以及侧屈(均为主动和被动)。通过组内相关系数(ICC)分析可靠性,并通过一致性界限以及5度、10度和15度内配对观察的百分比分析一致性。
对于无症状受试者,ICC在0.57至0.85之间,一致性界限在14.5度至27.0度之间。5度、10度和15度内配对观察的百分比分别为31%至57%、58%至90%和78%至93%。对于有症状受试者,ICC在0.36至0.91之间,一致性界限在9.6度至37.8度之间。5度、10度和15度内配对观察的百分比分别为17%至60%、33%至93%和50%至97%。
尽管使用了标准化方案和精密的测量系统,但颈部活动度的观察者间可靠性在质量上存在差异,中立位旋转、屈伸和侧屈的可靠性良好。