[Methods for automatic correction of the effect of various factors on the results of capnometry].

作者信息

Rozenblat L Sh, Sterlin Iu G, Simakhin A V

出版信息

Med Tekh. 2005 Jul-Aug(4):29-36.

Abstract

Factors affecting the results of carbon dioxide (CO2) concentration measurements and methods of their automatic compensation are considered for capnometers of two types: lateral flow (LF) and main flow (MF) capnometers. Advantages of the use of one-channel photometer in LF capnometer with automatic calibration were demonstrated. Compensation for temperature, pressure, and presence of concomitant gases and vapors in inhalation mixture is implemented automatically by introducing correction factors in CO2 concentration value. Calibration curves could be obtained empirically for specific capnometer model integrated into the gas analyzer. Carbon dioxide concentration values measured with LF and MF capnometers should be reduced to standard physical conditions STPD and BTPS conditions, respectively (body temperature and pressure, saturation with water vapors).

摘要

针对两种类型的二氧化碳(CO₂)浓度测量仪:旁流(LF)和主流(MF)二氧化碳测量仪,考虑了影响测量结果的因素及其自动补偿方法。展示了在具有自动校准功能的LF二氧化碳测量仪中使用单通道光度计的优势。通过在CO₂浓度值中引入校正因子,可自动实现对温度、压力以及吸入混合物中伴随气体和蒸汽的补偿。对于集成到气体分析仪中的特定二氧化碳测量仪型号,可通过实验获得校准曲线。使用LF和MF二氧化碳测量仪测得的二氧化碳浓度值应分别换算到标准物理条件下的STPD和BTPS条件(体温和压力,水蒸气饱和)。