Tri Auto ZX在根管再治疗过程中定位根尖孔能力的体外评估。

Ex vivo evaluation of the capacity of the Tri Auto ZX to locate the apical foramen during root canal retreatment.

作者信息

Alves A M H, Felippe M C S, Felippe W T, Rocha M J C

机构信息

School of Dentistry, Federal University of Santa Catarina, Florianopolis, Brazil.

出版信息

Int Endod J. 2005 Oct;38(10):718-24. doi: 10.1111/j.1365-2591.2005.01007.x.

Abstract

AIM

To evaluate ex vivo the capacity of the Tri Auto ZX to locate the apical foramen during root canal retreatment.

METHODOLOGY

The root canals of 62 maxillary and mandibular canines were prepared to a length 1 mm short of the apical foramen, to an apical size 35 using 1% sodium hypochlorite as an irrigant. Once prepared, the length of each tooth was measured directly using a size 15 K-Flexofile introduced in the canal until the tip was visible at the apical foramen. After the file was removed, its length was recorded to a precision of 0.01 mm using a calliper. These direct lengths (DL) became the 'gold standard' for comparison with the electronic lengths (EL) derived from the Tri Auto ZX. After direct measurement, the tooth was measured electronically (EL1) and the canals were filled using lateral condensation of gutta-percha cones and sealer. Seven days later the root filling was removed using solvent, Gates-Glidden burs, and K-files, and new electronic lengths determined (EL2). The electronic lengths (EL1 and EL2) were compared with the DL, and the differences were analysed statistically using the proportions test and Student's t-test.

RESULTS

At a tolerance limit of +/-0.5 mm, EL1 coincided with the DL in 76% of cases. Lengths obtained in the presence of remnant of filling material (EL2) coincided in 81% of cases. The proportions test used to compare these percentages showed no statistically significant difference between EL1 and EL2 (P > 0.05). The Student's t-test revealed a statistically significant difference (P < 0.05) between the means of the differences between DL/EL1 (-0.36 mm) and DL/EL2 (-0.04 mm).

CONCLUSION

The Tri Auto ZX was accurate to +/-0.5 mm in more than 80% of teeth following removal of root fillings.

摘要

目的

体外评估Tri Auto ZX在根管再治疗过程中定位根尖孔的能力。

方法

62颗上颌和下颌尖牙的根管预备至距根尖孔短1mm处,使用1%次氯酸钠作为冲洗剂,预备至根尖尺寸35。预备完成后,使用一根15号K-Flex锉插入根管直至其尖端在根尖孔处可见,直接测量每颗牙齿的长度。取出锉后,使用卡尺将其长度精确记录至0.01mm。这些直接长度(DL)成为与Tri Auto ZX得出的电子长度(EL)进行比较的“金标准”。直接测量后,对牙齿进行电子测量(EL1),并使用牙胶尖侧向加压和封闭剂充填根管。7天后,使用溶剂、Gates-Glidden钻和K锉去除根管充填物,并确定新的电子长度(EL2)。将电子长度(EL1和EL2)与DL进行比较,并使用比例检验和学生t检验对差异进行统计学分析。

结果

在±0.5mm的公差范围内,76%的病例中EL1与DL一致。在存在充填材料残留的情况下获得的长度(EL2)在81%的病例中一致。用于比较这些百分比的比例检验显示EL1和EL2之间无统计学显著差异(P>0.05)。学生t检验显示DL/EL1(-0.36mm)和DL/EL2(-0.04mm)之间差异的均值存在统计学显著差异(P<0.05)。

结论

去除根管充填物后,超过80%的牙齿中Tri Auto ZX的精度可达±0.5mm。