Gorny K R, Tradup D J, Hangiandreou N J
Department of Radiology, Mayo Clinic, Rochester, Minnesota 55905, USA.
Med Phys. 2005 Aug;32(8):2615-28. doi: 10.1118/1.1951095.
Systematic measurements of maximum depth of penetration (DOP) of ultrasound (US) scanners are essential for quality control (QC). Conventionally, DOP measurements are performed visually and as such they could be affected by various external factors, scanner control settings, and operator related errors. Automated methods should be free of the issues associated with interoperator dependence and are an attractive alternative to the visual DOP measurements. We implement and test three automated methods for measuring DOP. The methods base their measurements on signal to noise (SNR) analysis of uniform US phantom images. Two of the methods use pairs of phantom images. The third one uses a single phantom image and an "in-air" image. The validation tests included precision, sensitivity, repeatability, and usability in routine QC application. Methods based on pairs of phantom images measure the DOP with precision +/-0.2 cm or better. Precision of the single phantom image method is +/-0.05 cm, and that method is also the most sensitive of the three. All three methods are demonstrated to be repeatable among different users. Since the images for the DOP computation are collected free-hand the sensitivity to hand-transducer motion during image acquisition was also tested. Unlike the single-phantom-image based method, the methods using image pairs were found to be very sensitive to transducer motion and therefore less convenient for clinical QC applications. In conclusion, the single-phantom-image method is best suited for routine QC in a real-life clinical practice.
对超声(US)扫描仪的最大穿透深度(DOP)进行系统测量对于质量控制(QC)至关重要。传统上,DOP测量是通过目视进行的,因此可能会受到各种外部因素、扫描仪控制设置以及与操作员相关的误差的影响。自动化方法应避免与操作员间依赖性相关的问题,是目视DOP测量的一个有吸引力的替代方法。我们实现并测试了三种测量DOP的自动化方法。这些方法基于对均匀US体模图像的信噪比(SNR)分析进行测量。其中两种方法使用成对的体模图像。第三种方法使用单个体模图像和一个“空气中”的图像。验证测试包括在常规QC应用中的精度、灵敏度、可重复性和可用性。基于成对体模图像的方法测量DOP的精度为±0.2 cm或更高。单一体模图像方法的精度为±0.05 cm,并且该方法也是三种方法中最灵敏的。所有三种方法在不同用户之间都具有可重复性。由于用于DOP计算的图像是徒手采集的所以还测试了图像采集过程中对手持换能器运动的灵敏度。与基于单一体模图像的方法不同,发现使用图像对的方法对换能器运动非常敏感,因此在临床QC应用中不太方便。总之,单一体模图像方法最适合实际临床实践中的常规QC。