Huang Zhuo, Timerbaev Andrei R, Keppler Bernhard K, Hirokawa Takeshi
Department of Applied Chemistry, Graduate School of Engineering, Hiroshima University, 1-4-1 Kagamiyama, Higashi-hiroshima 739-8527, Japan.
J Chromatogr A. 2006 Feb 17;1106(1-2):75-9. doi: 10.1016/j.chroma.2005.09.042. Epub 2005 Oct 3.
Cisplatin is for a long time in clinical use as efficient antitumor drug. The success in cisplatin-based chemotherapy, however, strongly depends on how careful the drug's dosages are monitored in order to reduce severe side-effects and overcome cellular resistance. The use of micellar electrokinetic chromatography with direct UV detection is described for the determination of intact cisplatin in human serum. The main product of drug's hydrolytic metabolism, cis-diammineaquachloroplatinum(II), was quantified using capillary zone electrophoresis in combination with indirect UV detection and on-line transient isotachophoresis preconcentration. The detection limits of platinum species studied were about 2-3 micromol l(-1) that allows the proposed methods to be applied for quantification of the administered levels of cisplatin as well as drug's active metabolite. Changes in the speciation of cisplatin occurring after intravenous administration can also be monitored using these simple and convenient CE techniques.
顺铂作为一种高效的抗肿瘤药物已在临床使用很长时间。然而,基于顺铂的化疗的成功与否在很大程度上取决于对药物剂量的监测有多仔细,以减少严重的副作用并克服细胞耐药性。本文描述了使用胶束电动色谱结合直接紫外检测法测定人血清中完整顺铂的方法。药物水解代谢的主要产物顺二氨一水一氯铂(II),采用毛细管区带电泳结合间接紫外检测和在线瞬态等速电泳预富集进行定量分析。所研究的铂类物质的检测限约为2 - 3 μmol·L⁻¹,这使得所提出的方法可用于定量顺铂的给药水平以及药物的活性代谢物。使用这些简单便捷的毛细管电泳技术还可以监测静脉给药后顺铂形态的变化。