Suppr超能文献

使用Symmetry主动脉连接系统制作的隐静脉移植物近端吻合口的早期和中期通畅情况。

Early and midterm patency of the proximal anastomoses of saphenous vein grafts made with a Symmetry Aortic Connector System.

作者信息

Kitamura Hideki, Okabayashi Hitoshi, Hanyu Michiya, Soga Yoshiharu, Nomoto Takuya, Johno Hiroyuki, Nakano Jota, Matsuo Takehiko, Kai Masashi, Umehara Eitaro

机构信息

Department of Cardiovascular Surgery, Kokura Memorial Hospital, Kitakyusyu City, Fukuoka, Japan.

出版信息

J Thorac Cardiovasc Surg. 2005 Oct;130(4):1028-31. doi: 10.1016/j.jtcvs.2005.05.026.

Abstract

OBJECTIVE

The purpose of this study was to investigate (1) the early and midterm patency rates in saphenous vein grafts that were anastomosed with the Symmetry Aortic Connector System (St Jude Medical, Inc, St Paul, Minn) and (2) risk factors for graft occlusion.

METHODS

Thirty-one patients underwent off-pump coronary artery bypass grafting for proximal saphenous vein graft anastomoses with the aortic connector system. Intraoperative graft flow was studied with transit time flowmetry, and angiography was performed before discharge in 29 cases. Midterm (at least 1 year after the operation) saphenous vein graft patency was assessed by coronary angiography or 3-dimensional coronary computed tomography in 27 cases.

RESULTS

Postoperative angiography demonstrated a 100% patency rate and no significant stenosis at the connector-anastomosed sites. The 1-year patency rate of the saphenous vein grafts with the aortic connector system was 92.6%, with 2 cases of saphenous vein graft occlusion. Both of these cases had low graft flow and poor left ventricular function documented during the operation.

CONCLUSION

Once surgical errors had been ruled out, the 1-year patency rate of proximal saphenous vein grafts anastomosed with the aortic connector system was favorable. Poor left ventricular function and low graft flow during the operation were risk factors for midterm graft occlusion. Selection of the target vessel and its runoff may also be an important determinant of long-term patency.

摘要

目的

本研究旨在调查(1)使用Symmetry主动脉连接系统(圣犹达医疗公司,明尼苏达州圣保罗)吻合的大隐静脉移植物的早期和中期通畅率,以及(2)移植物闭塞的危险因素。

方法

31例患者接受了非体外循环冠状动脉旁路移植术,使用主动脉连接系统进行近端大隐静脉移植物吻合。术中采用渡越时间血流仪研究移植物血流,并对29例患者在出院前进行血管造影。27例患者在中期(术后至少1年)通过冠状动脉造影或三维冠状动脉计算机断层扫描评估大隐静脉移植物的通畅情况。

结果

术后血管造影显示通畅率为100%,连接吻合部位无明显狭窄。使用主动脉连接系统的大隐静脉移植物1年通畅率为92.6%,有2例大隐静脉移植物闭塞。这2例患者在手术期间均记录有移植物血流低和左心室功能差的情况。

结论

排除手术失误后,使用主动脉连接系统吻合的近端大隐静脉移植物1年通畅率良好。手术期间左心室功能差和移植物血流低是中期移植物闭塞的危险因素。目标血管及其血流情况的选择也可能是长期通畅的重要决定因素。