Kyele Nicholas R, Decanniere Klaas, van Silfhout Roelof G
School of Electrical and Electronic Engineering, The University of Manchester, PO Box 88, Manchester M60 1QD, UK.
J Synchrotron Radiat. 2005 Nov;12(Pt 6):800-6. doi: 10.1107/S0909049505031250. Epub 2005 Oct 18.
A compact, inexpensive and easy-to-construct two-dimensional in situ beam-position and profile monitor for synchrotron X-ray beamlines is presented. The device is based on the collection of spatially resolved scattered radiation from a polyimide foil. The X-ray beam passes through a foil placed in the path of the beam, which absorbs no more than 3% of the beam at 12 keV. The scattered radiation is collected at an angle of 90 degrees through a collimator located below the foil onto a CCD sensor. The device was tested on bending-magnet beamline BM26 at the ESRF synchrotron radiation source and has a positional sensitivity better than 10 microm with a large working range of 25 mm x 25 mm. Although the device is optimized for use in the range 10-12 keV, it can easily be modified for use with higher-energy beams by using a suitably chosen scattering foil.
介绍了一种用于同步加速器X射线束线的紧凑、廉价且易于构建的二维原位束流位置和轮廓监测器。该装置基于收集来自聚酰亚胺箔的空间分辨散射辐射。X射线束穿过放置在束流路径中的箔,该箔在12keV时吸收的束流不超过3%。散射辐射通过位于箔下方的准直器以90度角收集到CCD传感器上。该装置在欧洲同步辐射装置(ESRF)同步辐射源的弯铁束线BM26上进行了测试,其位置灵敏度优于10微米,工作范围为25毫米×25毫米。尽管该装置针对10-12keV范围内的使用进行了优化,但通过使用适当选择的散射箔,可轻松对其进行修改以用于更高能量的束流。