标准自动视野计、HRT共焦扫描激光眼科显微镜和GDx VCC扫描激光偏振仪之间的关系。

Relationships between standard automated perimetry, HRT confocal scanning laser ophthalmoscopy, and GDx VCC scanning laser polarimetry.

作者信息

Reus Nicolaas J, Lemij Hans G

机构信息

Glaucoma Service, The Rotterdam Eye Hospital, Rotterdam, The Netherlands.

出版信息

Invest Ophthalmol Vis Sci. 2005 Nov;46(11):4182-8. doi: 10.1167/iovs.04-1029.

Abstract

PURPOSE

This study was designed to determine and compare the relationships between visual function measured with standard automated perimetry (SAP) and structure, either as neuroretinal rim area measured with confocal scanning laser ophthalmoscopy (CSLO), or as retinal nerve fiber layer thickness determined by scanning laser polarimetry with variable corneal compensation (SLP-VCC).

METHODS

Forty-six healthy subjects and 76 glaucoma patients were examined with SAP, with CSLO by means of the commercially available Heidelberg Retina Tomograph I (HRT), and with SLP-VCC by means of the commercially available GDx VCC. The relationships between SAP, expressed either in the typically used decibel scale or as number of abnormal points in the total deviation probability plot, and CSLO and between SAP and SLP-VCC were described with linear and logarithmic regression analysis for global data and six individual sectors. The relationship between measurements with CSLO and SLP-VCC was fit with linear regression analysis.

RESULTS

The relationships between SAP and CSLO and between SAP and SLP-VCC appeared curvilinear for all sectors except the temporal one between SAP and SLP-VCC. For CSLO, a logarithmic fit was significantly better than a linear one for the global data and in the superotemporal and inferonasal sectors. For SLP-VCC, a curvilinear fit was better for the global data and in the superotemporal, superonasal, and inferonasal sectors. CSLO data correlated linearly with SLP-VCC data in all sectors, except temporally.

CONCLUSIONS

CSLO and SLP-VCC showed a very similar curvilinear relationship with SAP. The observed curvilinear relationships confirm earlier reports that these imaging devices appear to detect glaucomatous loss earlier than SAP.

摘要

目的

本研究旨在确定并比较通过标准自动视野计(SAP)测量的视觉功能与结构之间的关系,结构方面,要么是通过共焦扫描激光眼科显微镜(CSLO)测量的神经视网膜边缘面积,要么是通过可变角膜补偿扫描激光偏振仪(SLP-VCC)测定的视网膜神经纤维层厚度。

方法

对46名健康受试者和76名青光眼患者进行了SAP检查,使用市售的海德堡视网膜断层扫描仪I(HRT)进行CSLO检查,并使用市售的GDx VCC进行SLP-VCC检查。对于整体数据和六个单独的扇形区域,采用线性和对数回归分析来描述以常用分贝尺度表示的SAP或总偏差概率图中异常点数量与CSLO之间以及SAP与SLP-VCC之间的关系。用线性回归分析拟合CSLO和SLP-VCC测量值之间的关系。

结果

除了SAP与SLP-VCC之间的颞侧扇形区域外,所有扇形区域中SAP与CSLO之间以及SAP与SLP-VCC之间的关系均呈曲线关系。对于CSLO,对于整体数据以及颞上和鼻下扇形区域,对数拟合明显优于线性拟合。对于SLP-VCC,对于整体数据以及颞上、鼻上和鼻下扇形区域,曲线拟合更好。除颞侧外,所有扇形区域中CSLO数据与SLP-VCC数据均呈线性相关。

结论

CSLO和SLP-VCC与SAP显示出非常相似的曲线关系。观察到的曲线关系证实了早期报告,即这些成像设备似乎比SAP更早检测到青光眼性损害。