Shah J, dos Santos I, Haemmerich D, Valvano J W
Department of Biomedical Engineering, University of Texas at Austin, Texas, USA.
Med Biol Eng Comput. 2005 Jul;43(4):522-7. doi: 10.1007/BF02344735.
The primary objective of the paper was to present the design and analysis of an instrument to measure the heat convection coefficient h on the endothelial surfaces of arteries and veins. An invasive thermistor probe was designed to be inserted through the vessel wall and positioned on the endothelial surface. Electrical power was supplied to the thermistor by a constant temperature anemometry circuit. Empirical calibrations were used to relate electrical measurements in the thermistor to the h at the endothelial surface. As the thermal processes are strongly dependent on baseline blood temperature, the instrument was calibrated at multiple temperatures to minimise this potentially significant source of error. Three different sizes of thermistor were evaluated to optimise accuracy and invasiveness, and the smallest thermistors provided the best results. The sensitivity to thermistor position was evaluated by testing the device at multiple locations, varying both depth of thermistor penetration and position along the vessel. Finally, the measurement accuracy of the instrument was determined for the range of h from 430 to 4200 W m(-2)K, and the average error of the reading was 4.9% for the smallest thermistor. Although the instrument was designed specifically for measurements in the portal vein to obtain useful data for current numerical modelling, the device can be used in any large vessel.
本文的主要目的是介绍一种用于测量动脉和静脉内皮表面热对流系数h的仪器的设计与分析。设计了一种侵入式热敏电阻探头,通过血管壁插入并定位在内皮表面。通过恒温风速测量电路向热敏电阻供电。采用经验校准方法将热敏电阻的电学测量值与内皮表面的h联系起来。由于热过程强烈依赖于基线血液温度,因此在多个温度下对仪器进行校准,以尽量减少这一潜在的重大误差来源。评估了三种不同尺寸的热敏电阻,以优化准确性和侵入性,最小尺寸的热敏电阻给出了最佳结果。通过在多个位置测试该装置来评估对热敏电阻位置的敏感性,改变热敏电阻的穿透深度和沿血管的位置。最后,确定了该仪器在430至4200 W m(-2)K的h范围内的测量精度,对于最小尺寸的热敏电阻,读数的平均误差为4.9%。尽管该仪器是专门为在门静脉中进行测量而设计的,以便为当前的数值建模获取有用数据,但该装置可用于任何大血管。