Suppr超能文献

肝脏多探针射频消融的有限元分析

Finite-element analysis of hepatic multiple probe radio-frequency ablation.

作者信息

Haemmerich Dieter, Tungjitkusolmun Supan, Staelin S Tyler, Lee Fred T, Mahvi David M, Webster John G

机构信息

Department of Surgery, University of Wisconsin, Madison 53792, USA.

出版信息

IEEE Trans Biomed Eng. 2002 Aug;49(8):836-42. doi: 10.1109/TBME.2002.800790.

Abstract

Radio-frequency (RF) ablation is an important means of treatment of nonresectable primary and metastatic liver tumors. RF ablation, unlike cryoablation (a method of tumor destruction that utilizes cold rather than heat), must be performed with a single probe placed serially. The ablation of any but the smallest tumor requires the use of multiple overlapping treatment zones. We evaluated the performance of a configuration incorporating two hooked probes (RITA model 30). The probes were lined up along the same axis in parallel 20 mm apart. Three different modes applied voltage to the probes. The first mode applied energy in monopolar mode (current flows from both probes to a dispersive electrode). The second mode applied the energy to the probes in bipolar mode (current flows from one probe to the other). The third method applied the energy sequentially in monopolar mode (in 2-s intervals switched between the probes). We used the finite-element method (FEM) and analyzed the electric potential profile and the temperature distribution at the end of simulation of a 12-min ablation. The alternating monopolar mode allowed precise independent control of the amount of energy deposited at each probe. The bipolar mode created the highest temperature in the area between the probes in the configuration we examined. The monopolar mode showed the worst performance since the two probes in close vicinity create a disadvantageous electric field configuration. We, thus, conclude that alternating monopolar RF ablation is superior to the other two methods.

摘要

射频(RF)消融是治疗不可切除的原发性和转移性肝肿瘤的重要手段。与冷冻消融(一种利用低温而非高温破坏肿瘤的方法)不同,射频消融必须使用单个探头依次进行。除了最小的肿瘤外,对任何肿瘤进行消融都需要使用多个重叠的治疗区域。我们评估了一种包含两个钩形探头(RITA 30型)的配置的性能。探头沿同一轴线平行排列,相距20毫米。三种不同模式向探头施加电压。第一种模式以单极模式施加能量(电流从两个探头流向分散电极)。第二种模式以双极模式向探头施加能量(电流从一个探头流向另一个探头)。第三种方法以单极模式依次施加能量(在探头之间以2秒的间隔切换)。我们使用有限元方法(FEM),并在模拟12分钟消融结束时分析了电势分布和温度分布。交替单极模式允许精确独立控制每个探头处沉积的能量量。在我们研究的配置中,双极模式在探头之间的区域产生了最高温度。单极模式表现最差,因为附近的两个探头形成了不利的电场配置。因此,我们得出结论,交替单极射频消融优于其他两种方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验