Liu Lihong, Chen Xingguo, Hu Zhide
Department of Chemistry, Lanzhou University, Lanzhou, PR China.
Electrophoresis. 2005 Dec;26(23):4456-64. doi: 10.1002/elps.200500371.
A novel, rapid, and continuous online concentration approach based on field-amplified sample injection for the analysis of fangchinoline and tetrandrine was developed in this paper by combination of flow injection-MEKC. The BGE used was a solution composed of 75 mM H3PO4-triethylamine-2.5% v/v polyoxyethylene sorbitan monolaurate-20% v/v methanol buffer (pH* 5.0). The analytes prepared in 50% v/v aqueous ethanol were used as the test analytes. Sample was injected electrokinetically between plugs of water. When the cations reached the boundary between the water plug and BGE, they slowed down and became concentrated. Thereafter, MEKC was initiated for the separation. This results in 6.8-8.9-fold improvement in concentration sensitivity relative to conventional CE methods. The separation could be achieved within 10 min and sample throughput rate can reach up to 50/h. The repeatability (defined as RSD) was 4.8, 4.4% with peak height evaluation and 3.6, 0.94% with peak area evaluation for TET and FAN, respectively.