Aki C, Yilmaz S, Dilgin Y, Yagmur S, Suren E
Department of Molecular Biology, Faculty of Arts and Sciences, University of Canakkale Onsekiz Mart, Canakkale, Turkey.
Pharmazie. 2005 Oct;60(10):747-50.
The electrochemical oxidation and determination of natamycin has been carried out at a carbon paste electrode in aqueous solutions in the pH range of 2.5-10.30 by cyclic and differential pulse voltammetry. Best results were obtained with the differential pulse voltammetric technique in 0.5 M sulfuric acid at pH 1.82. The diffusion controlled nature of the waves was established. A differential pulse voltammetric technique for the determination of natamycin 0.5 M sulfuric acid which allows quantitation over the range of 2 x 10(-6)-8 x 10(-5) M range method is proposed. Limit of detection and limit of quantitatification were 1.5 x 10(-6) and 5 x 10(-6) M, respectively. Based on this study a simple, rapid, selective and sensitive voltammetric method was developed for the determination of natamycin in capsules. In order to validate the proposed method, UV spectroscopy was applied.