存在辐射阻尼时的反转恢复测量及其在磁共振中评估造影剂方面的意义。

Inversion recovery measurements in the presence of radiation damping and implications for evaluating contrast agents in magnetic resonance.

作者信息

Eykyn Thomas R, Payne Geoffrey S, Leach Martin O

机构信息

Cancer Research UK Clinical Magnetic Resonance Research Group, The Institute of Cancer Research, Royal Marsden Foundation NHS, Downs Road Sutton, Surrey, SM2 5PT, UK.

出版信息

Phys Med Biol. 2005 Nov 21;50(22):N371-6. doi: 10.1088/0031-9155/50/22/N03. Epub 2005 Nov 1.

Abstract

Relaxation measurements performed at high magnetic field in magnetic resonance (MR) may be adversely affected by the influence of radiation damping in concentrated samples such as water. We consider how the measured value of T1 is affected by this phenomenon for a gadolinium-doped water sample and for an undoped water sample and consider the implications for evaluating contrast agents. A simple method involving the application of a pulsed field gradient to de-phase residual transverse components of the magnetization is shown to be an effective method for suppressing this effect. Given the central role that measurement of the T1 of water plays in the assessment of contrast agents as well as a host of other MR applications, care should always be employed when measuring and interpreting T1 measurements at high magnetic fields.

摘要

在磁共振(MR)中,在高磁场下进行的弛豫测量可能会受到诸如水等浓缩样品中辐射阻尼影响的不利影响。我们考虑了钆掺杂水样和未掺杂水样中这种现象如何影响T1的测量值,并考虑了其对评估造影剂的影响。一种简单的方法,即应用脉冲场梯度使磁化的残余横向分量失相,被证明是抑制这种效应的有效方法。鉴于水的T1测量在造影剂评估以及许多其他MR应用中所起的核心作用,在高磁场下测量和解释T1测量值时应始终谨慎。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验