Sawada Akira, Yoda Kiyoshi, Numano Masumi, Futami Yasuyuki, Yamashita Haruo, Murayama Shigeyuki, Tsugami Hironobu
Advanced Technology R&D Center, Mitsubishi Electric Corporation, 8-1-1 Tsukaguchi Hommachi Amagasaki 661-8661, Japan.
Med Phys. 2005 Oct;32(10):3106-11. doi: 10.1118/1.2042247.
A new technique based on normalized binary image correlation between two edge images has been proposed for positioning proton-beam radiotherapy patients. A Canny edge detector was used to extract two edge images from a reference x-ray image and a test x-ray image of a patient before positioning. While translating and rotating the edged test image, the absolute value of the normalized binary image correlation between the two edge images is iteratively maximized. Each time before rotation, dilation is applied to the edged test image to avoid a steep reduction of the image correlation. To evaluate robustness of the proposed method, a simulation has been carried out using 240 simulated edged head front-view images extracted from a reference image by varying parameters of the Canny algorithm with a given range of rotation angles and translation amounts in x and y directions. It was shown that resulting registration errors have an accuracy of one pixel in x and y directions and zero degrees in rotation, even when the number of edge pixels significantly differs between the edged reference image and the edged simulation image. Subsequently, positioning experiments using several sets of head, lung, and hip data have been performed. We have observed that the differences of translation and rotation between manual positioning and the proposed method were within one pixel in translation and one degree in rotation. From the results of the validation study, it can be concluded that a significant reduction in workload for the physicians and technicians can be achieved with this method.
一种基于两个边缘图像之间归一化二值图像相关性的新技术已被提出用于质子束放射治疗患者的定位。使用Canny边缘检测器从患者定位前的参考X射线图像和测试X射线图像中提取两个边缘图像。在平移和旋转带边缘的测试图像时,两个边缘图像之间归一化二值图像相关性的绝对值会迭代最大化。每次旋转前,对带边缘的测试图像进行膨胀处理,以避免图像相关性急剧下降。为了评估所提方法的稳健性,通过在给定的旋转角度范围以及x和y方向的平移量范围内改变Canny算法的参数,从参考图像中提取了240个模拟的头部前视图边缘图像进行模拟。结果表明,即使带边缘的参考图像和带边缘的模拟图像之间的边缘像素数量显著不同,最终的配准误差在x和y方向上的精度为一个像素,旋转角度为零度。随后,使用几组头部、肺部和髋部数据进行了定位实验。我们观察到,手动定位和所提方法之间的平移和旋转差异在平移方面为一个像素以内,旋转方面为一度以内。从验证研究的结果可以得出结论,使用该方法可以显著减少医生和技术人员 的工作量。