Suppr超能文献

Characterization of a new certified reference material for human cardiac troponin I.

作者信息

Bunk David M, Welch Michael J

机构信息

Analytical Chemistry Division, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.

出版信息

Clin Chem. 2006 Feb;52(2):212-9. doi: 10.1373/clinchem.2005.051359. Epub 2005 Nov 23.

Abstract

BACKGROUND

To address the continuing need for the standardization of clinical human cardiac troponin I (cTnI) assays, NIST, with the assistance of the AACC/IFCC Cardiac Troponin I Standardization Committee, has developed a new certified reference material, Standard Reference Material (SRM) 2921: Human Cardiac Troponin Complex.

METHODS

The concentration of cTnI in SRM 2921 was determined through a combination of reversed-phase liquid chromatography (LC) with ultraviolet detection and amino acid analysis. Characterization of the intact troponin subunits was accomplished through reversed-phase LC coupled with mass spectrometry. Posttranslational modifications to the cTnI in SRM 2921 were investigated by combining proteolytic digestion with matrix-assisted laser desorption/ionization mass spectrometry. Additionally, reference concentration values for cTnT and cTnC were also determined.

RESULTS

The concentration of human cTnI in SRM 2921 is 31.2 (1.4) mg/L (where 1.4 mg/L is the uncertainty at a 95% level of confidence), as certified through a method that provides traceability to the International System of Units (SI). Reference concentration values of the cTnT and cTnC subunits were determined to be 36.9 (3.8) mg/L and 24.2 (1.3) mg/L, respectively.

CONCLUSIONS

This first cTnI reference material should provide SI traceability to clinical cTnI assays once commutability has been validated, and could assist in the international harmonization of cTnI assays as a tool for understanding the underlying causes of interassay variability.